Biomechanical Evaluation of an Endplate-Conformed Polycaprolactone-Hydroxyapatite Intervertebral Fusion Graft and Its Comparison With a Typical Nonconformed Cortical Graft

Author:

Agarwal Aakash1,Palepu Vivek1,Agarwal Anand K.1,Goel Vijay K.1,Yildirim Eda D.2

Affiliation:

1. e-mail:

2. e-mail:  Engineering Center for Orthopaedic Research Excellence (E-CORE), Departments of Bioengineering and Orthopaedic Surgery, Colleges of Engineering and Medicine, University of Toledo, Toledo, OH 43606

Abstract

In the thoracolumbar region, between 7% and 30% of spinal fusion failures are at risk for pseudarthrosis. From a biomechanical perspective, the nonconformity of the intervertebral graft to the endplate surface could contribute to pseudarthrosis, given suboptimal stress distributions. The objective of this study was to quantify the effect of endplate-graft conformation on endplate stress distribution, maximum Von Mises stress development, and stability. The study design used an experimentally validated finite element (FE) model of the L4–L5 functional spinal unit to simulate two types of interbody grafts (cortical bone and polycaprolactone (PCL)-hydroxyapatite (HA) graft), with and without endplate-conformed surfaces. Two case studies were completed. In Case Study I, the endplate-conformed grafts and nonconformed grafts were compared under without posterior instrumentation condition, while in Case Study II, the endplate-conformed and nonconformed grafts were compared with posterior instrumentation. In both case studies, the results suggested that the increased endplate-graft conformity reduced the maximum stress on the endplate, created uniform stress distribution on endplate surfaces, and reduced the range of motion of L4–L5 segments by increasing the contact surface area between the graft and the endplate. The stress distributions in the endplate suggest that the load sharing is greater with the endplate-conformed PCL-HA graft, which might reduce the graft subsidence possibility.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference27 articles.

1. Lumbar Degenerative Disk Disease;Radiology,2007

2. Estimating Cost of Care for Patients With Acute Low Back Pain: A Retrospective Review of Patient Records;J. Am. Osteopath. Assoc.,2009

3. Disc Degeneration: Current Surgical Options;Eur. Cell. Mater.,2010

4. The Tissue Origin of Low Back Pain and Sciatica: A Report of Pain Response to Tissue Stimulation During Operations on the Lumbar Spine Using Local Anesthesia;Orthop. Clin. North Am.,1991

5. United States Trends in Lumbar Fusion Surgery for Degenerative Conditions;Spine,2005

Cited by 22 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3