Effect of Flow Baffles on the Dialysate Flow Distribution of Hollow-Fiber Hemodialyzers: A Nonintrusive Experimental Study Using MRI

Author:

Poh Churn K.1,Hardy Peter A.2,Liao Zhijie1,Huang Zhongping1,Clark William R.3,Gao Dayong14

Affiliation:

1. Department of Mechanical Engineering, University of Kentucky, Lexington, KY

2. Center for Biomedical Engineering, University of Kentucky, Lexington, KY

3. Rental Division, Baxter Healthcare Corporation, McGaw Park, IL

4. and Center for Biomedical Engineering, University of Kentucky, Lexington, KY

Abstract

We used an innovative, nonintrusive MRI technique called the two-dimensional (2D) Phase-Contrast (2DPC) velocity-imaging technique to investigate the effect of flow baffles on the dialysate-side flow distribution in two different hollow-fiber hemodialyzers (A and B); each with flow rates between 200 and 1000 mL/min (3.33×10−6 and 1.67×10−5 m3/s). Our experimental results show that (1) the dialysate-side flow distribution was nonuniform with channeling flow occurred at the peripheral cross section of these hollow-fiber hemodialyzers, and (2) the existing designs of flow baffles failed to promote uniform dialysate-side flow distribution for all flow rates studies.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference32 articles.

1. Agishi, T., Ota, K., and Nose, Y., 1975, “Is Hollow Fibre Occlusion Due to Maldistributions of Blood?” Proc. Eur. Dial Transplant Assoc., 12, pp. 519–525.

2. Noda, I., Brown-West, D. G., and Gryte, C. C., 1979, “Effect of Flow Maldistribution on Hollow Fiber Dialysis—Experimental Studies,” J. Membr. Sci., 5, pp. 209–225.

3. Takesawa, S., Terasawa, M., Sakagami, M., Kobayashi, T., Hidai, H., and Sakai, K., 1988, “Nondestructive Evaluation by X-Ray Computed Tomography of Dialysate Flow Patterns in Capillary Dialyzers,” ASAIO Trans., 34, pp. 794–799.

4. Park, J. K., and Chang, H. N., 1986, “Flow Distribution in the Fiber Lumen Side of a Hollow-Fiber Module,” AIChE J., 32, pp. 1937–1947.

5. Zhang, J., Parker, D. L., and Leypoldt, J. K., 1995, “Flow Distributions in Hollow Fiber Hemodialyzers Using Magnetic Resonance Fourier Velocity Imaging,” ASAIO J., 41, pp. 678–682.