Affiliation:
1. Radioimmunoassay Unit, Stobhill General Hospital, Glasgow, G21 3UW
Abstract
Radioimmunoassay standard curves derived by four different automated calculation methods are compared with those derived manually for five assays of clinical importance (thyroxine, triiodothyronine, thyrotrophin, alphafetoprotein, and human growth hormone). The three curve fitting methods (linear or cubic regression on log x versus logit y, cubic regression on log x versus y) produced considerable distortion of the manually derived curves, which in some cases could have impaired the accuracy of an estimate in a clinical sample and altered the clinical interpretation. Distortion patterns varied with the particular antigen assayed and with experimental conditions. In contrast, a simple linear interpolation method produced little distortion and gave results which were close to those derived manually in each assay examined.
Subject
Clinical Biochemistry,General Medicine
Reference14 articles.
1. Semi-automated radioimmunoassays for total serum thyroxine and triiodothyronine
2. BASIC PRINCIPLES AND THEORY
3. Frazier G. R., Rodbard D. Radioimmunoassay data processing, second edition. Reports NIH-RIA 72–1 and NIH-RIA 72–2. National Technical Information Service, Springfield, Va., U.S.A. (1973).