Abstract

Introduction Patients undergoing cardiac surgery can experience significant thermal changes during the perioperative period and, for that reason, it is essential to monitor temperatures with adequate accuracy and precision during cardiopulmonary bypass (CPB). The primary aim of the current study was to measure the discrepancies between temperatures at different body sites during normothermic or mild hypothermic CPB. Methods 48 patients undergoing cardiac surgery participated in our study. Simultaneous temperatures were measured at nasopharynx, pulmonary artery, arterial outlet, venous inlet, forehead using a heat flux sensor, and urinary bladder at 5-min intervals throughout surgery. The Bland-Altman plot for repeated measures was used to assess concordance between methods. Results The duration of surgery was 360 min (interquartile range (IQR) 300–412), while the median cross-clamp time was 135 min (IQR 101–169). During the CPB time, the average difference between arterial outlet and nasopharyngeal temperature was −0.16°C (95% limits of agreement of ±0.93). The bias between arterial outlet and the venous inflow was 0.16°C and the 95% limits of agreement were −0.63 to 0.95°C. The Bland-Altman analysis showed an average difference between oxigenator arterial outlet and bladder probe of −0.62 (95% limits of agreement of ±1.3). The average difference between arterial outlet and Tcore™ temperatures was 0.08°C (95% limits of agreement of ±1.46). 25 patients (52.08%) presented nasopharyngeal temperatures higher than 37°C in the post-CPB period, but none of them exceeded 38°C. Conclusions Perfusionists should be cautious when using the nasopharyngeal site as the only surrogate of brain temperature, even in normothermic cardiac surgery because the precision of measurements is not entirely adequate.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3