Efficacy of Plasma free Hemoglobin for detecting centrifugal pump thrombosis

Author:

Kuroda Taiyo1ORCID,Mutsuga Masato2,Yamada Masao1,Yamakawa Masato1,Yuhara Satoshi1,Hasegawa Hiroki1,Yokote Jun1,Yokoyama Yukifusa1,Yamada Tetsuya3,Koyama Tomio2,Usui Akihiko2

Affiliation:

1. Department of Thoracic and Cardiovascular Surgery, Ogaki Municipal Hospital, Ogaki, Gifu, Japan

2. Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan

3. Department of Clinical Engineering, Ogaki Municipal Hospital, Ogaki, Gifu, Japan

Abstract

Introduction: Lactate dehydrogenase (LDH) is widely used as an indicator of pump thrombosis in a centrifugal pump. However, due to the low specificity of LDH, pump thrombosis is difficult to detect in the clinical environment. We measured plasma free hemoglobin (pfHb) with the portable device in ICU. The goal of this investigation is to evaluate its diagnostic ability for pump thrombosis. Methods: We enrolled 31 consecutive patients who needed Extracorporeal Membrane Oxygenation (ECMO) therapy and pfHb was determined with HemoCue® plasma/Low Hb photometer. Pump thrombosis was analyzed macroscopically at the timing of pump explantation or exchange. Also, we divided the pump thrombosis into a grading scale by the place of thrombosis. Results: The median of peak pfHb was significantly lower in the none thrombus group (0.03 g/dL) than that of in the thrombus group (0.05g/dL) ( p = 0.01). In our grading criteria, pfHb was significantly higher when the thrombus is existing near the shaft ( p = 0.015). Contrary, no significant difference was found for LDH. The ROC analysis of pfHb revealed an AUC of 0.77 for detecting pump thrombosis with the best statistical cutoff value at 0.05 g/dL (specificity, 78%; sensitivity, 77%). Also, ROC analysis of LDH was performed (AUC, 0.44; cutoff value, 1200 IU/L; specificity, 59%; sensitivity, 54%) and compared with pfHb. AUC was significantly higher in pfHb ( p = 0.04). Conclusion: Our results showed the efficacy of pfHb for detecting centrifugal pump thrombosis.

Publisher

SAGE Publications

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3