Affiliation:
1. Department of Neurosurgery, University of California Los Angeles, USA
2. Department of Interventional Neuroradiology, University of California Los Angeles, USA
Abstract
Due to technical limitations, small, distal, and tortuous intracranial pathology is sometimes out of reach of the current armamentarium of microcatheters designed for intracranial coil embolization. The Marathon microcatheter (Medtronic, Minneapolis, Minnesota, USA), designed specifically for the delivery of Onyx, is longer and more flexible than most coil delivery catheters. We report on nine patients (three with arteriovenous fistula, three with arteriovenous malformation, two with intracranial aneurysm, and one with tumor) where Marathon was used to deliver commercially available platinum coils. We also conducted laboratory compatibility testing and conclude that the Marathon can be used as a coil delivery catheter for Barricade coils (Blockade Medical, Irvine, California, USA) with diameter less than 0.012 in.
Cited by
22 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献