VMAT Optimization and Dose Calculation in the Presence of Metallic Hip Prostheses

Author:

Parenica Holly M.1ORCID,Mavroidis Panayiotis2,Jones William3,Swanson Gregory4,Papanikolaou Niko1,Stathakis Sotirios1

Affiliation:

1. Department of Radiation Oncology, School of Medicine, The University of Texas Health Science Center at San Antonio, TX, USA

2. Department of Radiation Oncology, School of Medicine, University of North Carolina, Chapel Hill, NC, USA

3. Department of Radiation Oncology, South Texas Veterans Health Care System, San Antonio, TX, USA

4. Department of Radiation Oncology, Baylor Scott & White Clinic-Temple, Temple, TX, USA

Abstract

Introduction:This research quantifies and compares the effect of hip prostheses on dose distributions calculated using collapsed cone convolution superposition and Monte Carlo (with and without correcting for the density of the implant and surrounding tissues). The use of full volumetric modulated arc therapy arcs versus volumetric modulated arc therapy arcs avoiding the hip implants (skip arcs) was also studied.Materials and Methods:Six prostate patients with hip prostheses were included in this study. The hip prostheses and the streaking artifacts on the computed tomography images were contoured by a single physician, and full volumetric modulated arc therapy arcs were created in the Pinnacle3 TPS. Copies of each plan were made, and the doses were recalculated with the densities of the prostheses and surrounding tissues overridden. The plans were then exported to Monaco and recalculated using a Monte Carlo dose calculation algorithm, with and without densities of the prosthesis and surrounding tissues overridden.Results:With density overrides, Pinnacle3 had a 4.4% error for ion chamber measurements. Monaco was within 0.2% of ion chamber measurement when density overrides were used. On average, when density overrides were used in Pinnacle3 for patient dose calculations, the planning target volume D95 value dropped from 99.3% to 82.7%. Monaco also showed decreased planning target volume coverage when plans were recalculated with correct density information. Full arc plans (with density overrides) for the patient with a bilateral prosthesis provided significant bladder sparing and some rectal sparing compared to skip arc plans.Conclusion:When planning for prostate patients with hip prostheses, correct density information for implants and surrounding tissues should be used to optimize the plan and ensure optimal accuracy. If available, a Monte Carlo algorithm should be used as a second check. Full arcs could be used to spare dose to organs at risk, while maintaining adequate planning target volume coverage, when using a Monte Carlo dose calculation algorithm.

Funder

Cancer Prevention and Research Institute of Texas

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3