BladderScan Feedback Method in Predicting Bladder Filling for Prostate Radiotherapy: A Prospective Study

Author:

Kuo Deng-Yu1ORCID,Hsu Chen-Yang23,Wang Wei-Chun2,Chen Hsiu-Hsi2,Shueng Pei-Wei14

Affiliation:

1. Division of Radiation Oncology, Far Eastern Memorial Hospital, New Taipei City, Taiwan

2. Institute of Epidemiology and Preventive Medicine, College of Public Health, National Taiwan University, Taipei

3. Dachung Hospital, Miaoli, Taiwan

4. Faculty of Medicine, School of Medicine, National Yang Ming Chiao Tung University, Taipei, Taiwan

Abstract

Purpose: Approximately 5%–10% of men who receive prostate cancer radiotherapy will suffer from radiation cystitis. Bladder filling before the administration of radiotherapy results in lower radiation exposure to the bladder. BladderScan, an ultrasound-based bladder volume scanner, has the potential to evaluate bladder volume during radiotherapy; thus, a prospective pilot study was initiated. Methods: Eleven men receiving tomotherapy for localized prostate cancer were enrolled. The validity of BladderScan was evaluated by comparing the measurements from BladderScan with the calculated volume from megavoltage computed tomography (MVCT). With a crossover design to compare different methods in bladder filling, the radiotherapy was divided into 2 sequences. Conventional method: the patient was asked to drink water after voiding urine. The amount of water and the duration of waiting were the same as in the setting of the simulation. BladderScan feedback method: the bladder filling procedure depended on the BladderScan measurements. Results: There were 314 sets of data from 11 patients. The correlation coefficient between VBS and VCT was 0.87, where VBS is the mean volume of 3 measurements by BladderScan and VCT is the bladder volume derived from MVCT. The BladderScan feedback method resulted in a significant larger bladder volume than the conventional method, with a mean difference of 36.9 mL. When the failure was defined as VCT <80% of planned volume, the BladderScan feedback method brought about a relative reduction in the failure rate with an odds ratio of 0.44 and an absolute reduction of 9.1%. Conclusion: The accuracy of BladderScan was validated by MVCT in our study. The BladderScan feedback method can help patients fill the bladder adequately, with a larger bladder volume and a lower failure rate.

Funder

Far Eastern Memorial Hospital

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3