Development and validation of a simple microwave-assisted digestion sample preparation technique for the estimation of aluminium and magnesium in a few pharmaceutical dosage forms by an inductively coupled-mass spectrometer (ICP-MS)

Author:

Pal Ashish Kumar1ORCID,Sundararajan Raja1

Affiliation:

1. Department of Pharmaceutical Analysis, GITAM Institute of Pharmacy, Gandhi Institute of Technology and Management (GITAM) (Deemed to be University), Visakhapatnam, AP, India

Abstract

Rationale: A simple, sensitive, reliable, validated, inductively coupled plasma mass spectrometric method for the determination of aluminium and magnesium using a simple common microwave-assisted digestion sample preparation technique for a few commonly used formulations was developed and validated according to International Conference on Harmonization Q3D and the United States Pharmacopeia general chapter <232> and <233>. The following pharmaceutical dosage forms were considered for estimation of aluminium and magnesium: Alumina, magnesia simethicone oral suspension, Alumina, magnesia simethicone chewable tablets, alumina and magnesia oral suspension, alumina and magnesium carbonate oral suspension. Methods: The methodology included optimizing a common microwave assisted digestion method, selecting the isotopes, choosing the measurement technique, and designating internal standards. The finalized microwave assisted procedure was a two-step program where in the first step the samples were ramped for 10 min to a temperature of 180 °C and hold for 5 min followed by ramping for 10 min to a temperature of 200 °C and hold for 10 min. Magnesium (24Mg) and aluminium (27Al) isotopes were finalized, internal standard assigned for both the isotopes was yttrium (89Y) with Helium (kinetic energy discrimination-KED) as the measuring mode. System suitability was run before initiating analysis to ensure that system performance was consistent. Results: Analytical validation parameters like specificity, linearity (from 25% to 200% of sample concentration), the detection limit and the limit of quantification were established. For all these dosage forms, the method's precision was demonstrated by analyzing the percentage relative standard deviation for six injections. Accuracy was established from 50% to 150% of instrument working concentration (J-levels) for aluminium and magnesium for all the formulations and was found to be within the range of 90–120%. Conclusion: This common analysis method, along with the common microwave-digestion technique applies to numerous types of matrices for a finished dosage form with aluminium and magnesium.

Publisher

SAGE Publications

Subject

Spectroscopy,Atomic and Molecular Physics, and Optics,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3