Abstract

The socket of a transtibial prosthesis is a structural part customized to a patient’s amputated residual lower limb. The free-form geometry of the socket can be suitable for additive manufacturing (AM) to save time and cost. However, the mechanical fracture of additively manufactured lower limb prostheses is not yet fully understood. A novel experimental method and numerical approach by finite element method (FEM) to test the strength and fracture behavior of a lower limb prosthetic socket of acrylonitrile butadiene styrene (ABS), reverse-engineered using computer-aided design (CAD) from the actual amputee’s residual limb and manufactured using fused filament fabrication (FFF) are proposed in the present work. The mechanical behavior, von Mises stress distribution, and the damage status of layered AM sockets of different thicknesses were simulated by FEM using Hashin’s transversely isotropic mechanical damage model, initially developed for composite materials. The experimental work showed that the fracture failure initiated at the corner of the lobe in the 4 mm thickness socket at a failure load of 918.5 N. The FEM results predicted this failure load to be 896.6 N, with only a 2.45% error as compared to the experiment. The failure loads predicted by FEM in the sockets with thicknesses of 3, 5, and 6 mm were 618.1, 1008.6, and 1105.2 N, respectively. The present work provides a dependable method for testing a below-knee prosthetic socket against static failure and arriving at a factor-of-safety (FoS) based socket thickness selection for any amputee.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3