Abstract

Ultrasound guidance for epidural block has improved clinical blind-trial problems but the design of present ultrasonic probes poses operating difficulty of ultrasound-guided catheterization, increasing the failure rate. The purpose of this study was to develop a novel ultrasonic probe to avoid needle contact with vertebral bone during epidural catheterization. The probe has a central circular passage for needle insertion. Two focused annular transducers are deployed around the passage for on-axis guidance. A 17-gauge insulated Tuohy needle containing the self-developed fiber-optic–modified stylet was inserted into the back of the anesthetized pig, in the lumbar region under the guidance of our ultrasonic probe. The inner transducer of the probe detected the shallow echo signals of the peak-peak amplitude of 2.8 V over L3 at the depth of 2.4 cm, and the amplitude was decreased to 0.8 V directly over the L3 to L4 interspace. The outer transducer could detect the echoes from the deeper bone at the depth of 4.5 cm, which did not appear for the inner transducer. The operator tilted the probe slightly in left-right and cranial-caudal directions until the echoes at the depth of 4.5 cm disappeared, and the epidural needle was inserted through the central passage of the probe. The needle was advanced and stopped when the epidural space was identified by optical technique. The needle passed without bone contact. Designs of the hollow probe for needle pass and dual transducers with different focal lengths for detection of shallow and deep vertebrae may benefit operation, bone/nonbone identification, and cost.

Funder

National Health Research Institutes

Ministry of Science and Technology, Taiwan

Publisher

SAGE Publications

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3