Inflammatory Response of a New Synthetic Dialyzer Membrane. a Randomised Cross-over Comparison between Polysulfone and Helixone

Author:

Stefoni S.1,Colì L.1,Cianciolo G.1,Donati G.1,Ruggeri G.1,Ramazzotti E.2,Pohlmeier R.3,Lang D.3

Affiliation:

1. Nephrology Dialysis and Renal Transplantation Unit, Department of Clinical Medicine and Applied Biotechnology, S. Orsola University Hospital, Bologna - Italy

2. Central Laboratory, S. Orsola University Hospital, Bologna - Italy

3. Fresenius Medical Care, Bad Homburg - Germany

Abstract

Hemodialysis patients suffer from chronic inflammation due to intradialytic contact of blood with artificial materials. The FX 60 dialyzer which belongs to the new FX-class series of dialyzers is composed of the new membrane Helixone®. This membrane is derived from the original Fresenius Polysulfone® membrane. The FX-class design is based on modified geometry of fibres and housing and has resulted in a new dialyzer with improved efficiency, safety and ease of handling compared to the F series (F 60S) dialyzer. The aim of the study was to investigate whether the biocompatibility pattern in terms of inflammatory parameters of the new type of polysulfone dialyzer has changed compared to the standard. A clinical in vivo study was conducted to compare the intradialytic inflammatory response of the two dialyzers, FX 60 and F 60S. Eight chronic dialysis patients were selected for the study: mean age 65.5±15.5 years, mean time on dialysis 100±95 months. The randomized cross-over study involved a treatment period of 2 weeks (total 6 sessions), one week with each dialyzer, starting with one or the other according to the randomization scheme. Blood samples were taken at 0 (T0), 15, 60, and 240 minutes to evaluate white blood cell (WBC) count, complement factor C5a, leukocyte elastase, soluble intercellular adhesion molecule 1 (sICAM-1), platelet count, C-reactive protein (CRP). At 15 min, WBC count showed a comparably, low decrease for both dialyzers: −7.6 % for FX 60 versus −6.6 % for F 60S, p=not significant (ns). At the same time the C5a concentration decreased from 15.0±7.5 ng/ml to 13.5±6.7 ng/ml (p=ns) for FX 60, and from 15.1±12.5 ng/ml to 14.9±25.0 ng/ml for F 60S (p=ns). The elastase concentration progressively increased over time with no statistical difference between the two dialyzers. The levels of sICAM-1, CRP, and platelet count were similar at each time point for both dialyzers, varying around the baseline values (p=ns). No significant difference emerged in terms of inflammatory response between the two dialyzers, demonstrating that the biocompatibility of the F-series was maintained in the FX-class series of dialyzers and is independent of design factors.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3