OPTIMIZING THE SPATIAL RESOLUTION OF MAMMOGRAPHY IMAGING QUALITY USING A CIRS-016A LINE PAIR GAUGE AND THE TAGUCHI METHODOLOGY

Author:

LIN YA-HUI12,SHEN CHENG-MAO134,TSENG YEN-LING15,PAN LUNG-KWANG1ORCID,PENG BING-RU1

Affiliation:

1. Department of Medical Imaging and Radiological Science, Central Taiwan University of Science and Technology, Takun, Taichung 406, Taiwan, ROC

2. Department of Clinical Pharmacy, Taichung Armed, Forces General Hospital, Taichung 406, Taiwan, ROC

3. Division of Gastroenterology and Hepatology, Department of Internal Medicine, Taichung Armed Forces, General Hospital, Taichung 411, Taiwan, ROC

4. Division of Gastroenterology and Hepatology, Department of Internal Medicine, Tri-Service General Hospital, National Defense Medical Center, Taipei 114, Taiwan, ROC

5. Department of Radiology, Lukang Christian Hospital, Lukang, Changhua 505, Taiwan, ROC

Abstract

This study optimized spatial resolution of mammography imaging quality using a CIRS-016A commercial line gauge and the Taguchi methodology. The line gauge with a precise line pair from 5[Formula: see text]lp/mm to 20[Formula: see text]lp/mm was placed on top of triangular PMMA plates to simulate the female breast undergoing mammography. Five factors: target/filter, kVp, mAs, PMMA plate thickness, and compression force, were organized into 18 groups according to the Taguchi L[Formula: see text] orthogonal array. Tactically, the 18 various combinations of factors could provide similar confidence levels, as those following the full factorial combination in reality. Seven experienced radiology experts judged the 18 imaging qualities based on contrast, sharpness, and spatial resolution. Then the signal-to-noise ratio was calculated according to the “the larger, the better” ranking order. The optimal preset of mammography was verified from the unique fish bone plot and the follow-up analysis of variance (ANOVA) test. The optimal combination of factors was as follows: Rh/Ag as target/filter, 32[Formula: see text]kVp, 36[Formula: see text]mAs, a 45[Formula: see text]mm thick PMMA plate, and a 13[Formula: see text]daN compression force in routine diagnosis. The concurrent resolution of 6[Formula: see text]lp/mm or about a 0.09[Formula: see text]mm minimum detectable difference (MDD) was superior to 5[Formula: see text]lp/mm of the conventional preset or combinations of factors of either highest Avg or lowest std. Compared to other studies with various facilities, this was the finest resolution among the routine X-ray, cardiac X-ray or computed tomography (CT), and computed tomography angiography (CTA).

Publisher

World Scientific Pub Co Pte Ltd

Subject

Biomedical Engineering