NUMERICAL ANALYSIS OF BALLOON EXPANDABLE STENT DEPLOYMENT INSIDE A PATIENT-SPECIFIC STENOTIC CORONARY ARTERY TO INVESTIGATE THE INSTANT MECHANICAL BEHAVIORS

Author:

XU CHUANGYE12,LIU XIUJIAN12,PAN LIANQIANG12,WU GUANGHUI12,SHU LIXIA12,HE YUNA12,MA LIPING3,LIN CHANGYAN12

Affiliation:

1. Department of Biomedical Engineering, Beijing Anzhen Hospital, Capital Medical University, Beijing 100029, P. R. China

2. Department of Biomedical Engineering, Beijing Institute of Heart, Lung, and Blood Vessel Diseases, Beijing 100029, P. R. China

3. General Practice, Beijing Anzhen Hospital, Capital Medical University, Beijing 100029, P. R. China

Abstract

The instant mechanical behaviors of stenotic coronary artery and deployed stents have significant impacts on percutaneous coronary intervention prognosis. However, they could not be obtained directly from the current examination techniques, which are commonly used in clinical practice. Thus, we intend to investigate the instantaneous mechanical behaviors of deployed stent and artery through virtually stenting technology based on a real clinical case in assessment of geometric and biomechanical characteristics. Method: Finite element analysis models, including rigid guide catheter, six-folded balloon with conical tip, crimped and bended stent, stenotic coronary artery with soft plaques, were simulated through virtual mechanical expansion and recoil procedure. The morphology changes of coronary lumen, strain and stress distribution of involved components at different stages and apposition of stent struts were analyzed. Results: Lumen in the stenotic region restored patency obviously at maximum expansion and had an elastic recoil about 13.5% later. The maximum principal stress distribution of artery walls and plaque was mainly concentrated in the stenotic segment with the peak value of 1.252[Formula: see text]MPa and 2.975[Formula: see text]MPa at max expansion, 0.713[Formula: see text]MPa and 1.25[Formula: see text]MPa after recoil, respectively. The higher von Mises stress and plastic equivalent strain of stent were present at the bended strut and inter-ring connectors with the peak value of 714.2[Formula: see text]MPa and 0.2385 at max expansion, 694[Formula: see text]MPa and 0.2276 after recoil. Slight malappositions were found in the proximal segment and struts distribution in the stenotic sites showed certain asymmetry. Conclusion: The instant mechanical behaviors of artery and stent could be evaluated through virtual stenting approach in assessment of geometric and biomechanical characteristics. This may contribute to choosing the best stenting schemes and predicting the clinical outcomes for a specific patient.

Publisher

World Scientific Pub Co Pte Lt

Subject

Biomedical Engineering

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. DESIGN AND ANALYSIS OF NEW STENT PATTERNS FOR ENHANCED PERFORMANCE;Journal of Mechanics in Medicine and Biology;2020-08

全球学者库

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"全球学者库"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前全球学者库共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2023 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3