Interference between Venus-A Valve and Anterior Mitral Valve Leaflets after Transcatheter Aortic Valve Replacement: Insight from FEops HEARTguide simulation

Author:

Wang Yong,Liu Ting,Zeng Ying,Debusschere Nic,Rocatello Giorgia,Cheng Sihang,Li Ping,Qian Dehui,Yu Shiyong,Jin Jun

Abstract

AbstractBackgroundScarce data exist regarding the occurrence of mitral valve interference after transcatheter aortic valve replacement (TAVR) with Venus-A valve implantation. Several case reports have noted that the anterior mitral leaflet (AML) is mechanically affected by the prosthesis frame, particularly when implanted in a low position. This study aimed to investigate the potential factors influencing the clinical outcomes of AML interference after Venus-A valve implantation.MethodsWe retrospectively included 20 severe aortic valve stenosis patients who had undergone TAVR and had been implanted with the Venus-A valve at our hospital between October 2020 and June 2021. Pre- and post-procedural CT scans were used for the FEops HEARTguide simulation. Anatomically influencing factors were measured using the 3mensio software and derived from the FEops HEARTguide. The prosthesis-AML interference (PAI) was defined when it met both of two criteria:1) significant interference and limited AML movement shown by transthoracic or transoesophageal echocardiography, and 2) more than half cell intersection between the simulated Venus-A valve and the reconstructed AML revealed by the FEops HEARTguide. Anatomical factors and clinical outcomes were compared between the PAI and non-PAI groups.ResultsNine PAI patients and 11 non-PAI cases were identified. PAI was associated with shorter mitral-aortic annulus distance (2.7±1.7 mm vs 5.0±2.2 mm,P= 0.019), larger prosthesis valve size (P= 0.013), deeper implantation (12.2±3.3 mm vs 6.2±2.9 mm at non-coronary cusp side,P< 0.001) and less calcification of non-coronary cusp (median calcification score, 52.2 mm3vs 156.0 mm3,P= 0.046). Regarding the clinical impact, PAI was associated with a higher rate of moderate or severe perivalvular leakage before discharge than those associated with the absence of PAI, with no difference in haemodynamic parameters and incidence of adverse events at the 30-day and 12-month follow-ups between the groups.ConclusionsInterference between the Venus-A prosthesis valve and AML after TAVR was associated with a shorter mitral-aortic annulus distance, larger prosthesis usage, greater implantation depth, and less calcification of the non-coronary cusp. However, further studies are required to explore its long-term clinical impact.

Publisher

Cold Spring Harbor Laboratory

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3