Affiliation:
1. United Kingdom Health Security Agency, RCE , Chilton OX11 0RQ , United Kingdom
Abstract
Abstract
Passive diffusion radon monitoring devices were exposed to a range of radon concentrations and their variations. The experimental results, backed up by the statistical analysis, showed that the radon concentration had virtually no effect on the final integrated exposure results, with the exception for the very low concentration. Therefore, it was proven that such devices are truly integrated exposure devices.
Publisher
Oxford University Press (OUP)
Reference10 articles.
1. Nuclear tracks today: strengths, weaknesses, challenges;Durani;Radiat. Measur.,2008
2. Results of the 2021 intercomparison of passive radon detectors, UKHSA RCE-002, GOV-13157;Miller,2022
3. Inter-comparison of radon detectors for one to four week measurement periods;Gunning;J. Radiol. Prot.,2016
4. An intercomparison of indoor radon data using NTD and different dynamic recording systems;Espinosa;Radiat. Measur.,2013
5. Ageing effects in CR39;Portwood;Int. J. Radiat. Appl. Inst. D: Nucl. Tracks Radiat. Measur.,1986
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献