Abstract
Abstract
The Hitachi L-8500A is a newly available apparatus for amino acid (AA)analysis that allows automatic on-line mixing of the ninhydrin reagent. The within-run precision (human plasma pools at three different concentrations) showed CVs <3.8% except for the lowest concentration of citrulline (4.4%), Tyr (4.5%), and α-aminobutyric acid (7.6%), and for the intermediate concentration of Asp (8.7%). Between-run precision (CV) was <3.1% for 17 AAs and <8.0% for 24 of 25 AAs (CV Asp = 12.0%). For retention times, within-run precision was <0.4% and between-run precision <1.8%. Excellent relations were found between the results from the Hitachi L-8500A and the widely used Beckman 6300 analyzer (0.929≤ r ≤0.999). The detection was still linear at 5 μmol/L except for Pro and hydroxyproline (20 μmol/L). The upper limit was at least 2500 μmol/L for 13 AAs and at least 1000 μmol/L for 27 of 29 AAs (anserine = 500, Val = 600 μmol/L). Values from 100 human plasma samples agreed with previously published data. We conclude that the results obtained with the Hitachi L-8500A are satisfactory when compared with those of other AA analyzers utilizing the same method. Furthermore, the Hitachi L-8500A displays several advantages including programming flexibility, microsample capacity, low noise plotting, ammonia filtering, and manual repacking of the analytical column.