ESTIMATION OF PATIENT RADIATION DOSES FROM MULTI-DETECTOR COMPUTED TOMOGRAPHY ANGIOGRAPHY PROCEDURES IN TANZANIA

Author:

Ngaile J E12ORCID,Msaki P K2,Suleiman S A1,Chuma F M1,Muhogora W E1,Jusabani A M3

Affiliation:

1. Radiation Control Directorate, Tanzania Atomic Energy Commission, P O Box 743 Arusha, Tanzania

2. Department of Physics, University of Dar es Salaam, P O Box 35063 Dar es Salaam, Tanzania

3. Department of Radiology, The Agha Khan Hospital, P O Box 2289 Dar es Salaam, Tanzania

Abstract

Abstract The aim of the present study was to estimate the volume CT dose index (CTDIvol), dose length product (DLP) and effective dose (ED) to patients from five multi-detector computed tomography angiography (MDCTA) procedures: brain, carotid, coronary, entire aorta and lower limb from four medical institutions in Tanzania; to compare these doses to those reported in the literature, and to compare the data obtained with ICRP 103 and Monte Carlo software. The radiation doses for 217 patients were estimated using patient demographics, patient-related exposure parameters, the geometry of examination and CT-Expo V 2.4 Monte Carlo-based software. The median values of the CTDIvol, DLP and ED for MDCTA procedures of the brain and carotids were 36.8 mGy, 1481.0 mGy∙cm and 5.2 mSv, and 15.9 mGy, 1224.0 mGy∙cm and 7.8 mSv, respectively; while for the coronary, entire aortic, and lower limbs were 49.4 mGy, 1493.0 mGy∙cm and 30.6 mSv; 16.2 mGy, 2287.0 mGy∙cm and 41.1 mSv; and 6.4 mGy, 1406.0 mGy∙cm and 10.5 mSv, respectively. The ratio of the maximum to minimum ED values to individual patients across the four medical centers were 41.4, 11.1, 4.6, 9.5 and 37.4, respectively, for the brain, carotid, coronary, entire aortic and lower limb CT angiography procedures. The mean values of CTDIvol, DLP and ED in the present study were typically higher than the values reported from Kenya, Korea and Saudi Arabia. The 75th percentile values of the DLP were above the preliminary diagnostic references levels proposed by Kenya, Switzerland and Korea. The observed wide range of examination scanning protocols and patient doses for similar MDCTA procedures within and across hospitals; and the observed relatively high patient doses compared to those reported in the literature, call for the need to standardize scanning protocols and optimise patient dose from MDCTA procedures.

Funder

Tanzania Commission for Science and Technology

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3