Calibration of an airborne HO<sub><i>x</i></sub> instrument using the All Pressure Altitude-based Calibrator for HO<sub><i>x</i></sub> Experimentation (APACHE)
-
Published:2020-05-27
Issue:5
Volume:13
Page:2711-2731
-
ISSN:1867-8548
-
Container-title:Atmospheric Measurement Techniques
-
language:en
-
Short-container-title:Atmos. Meas. Tech.
Author:
Marno DanielORCID, Ernest Cheryl, Hens Korbinian, Javed Umar, Klimach Thomas, Martinez Monica, Rudolf MarkusORCID, Lelieveld JosORCID, Harder HartwigORCID
Abstract
Abstract. Laser-induced fluorescence (LIF) is a widely used technique for
both laboratory-based and ambient atmospheric chemistry measurements.
However, LIF instruments require calibrations in order to translate
instrument response into concentrations of chemical species. Calibration of
LIF instruments measuring OH and HO2 (HOx) typically involves the photolysis of water vapor by 184.9 nm light, thereby producing quantitative amounts of OH and HO2. For ground-based HOx instruments, this method of calibration is done at one pressure (typically ambient pressure) at the instrument inlet. However, airborne HOx instruments can experience varying cell pressures, internal residence times, temperatures,
and humidity during flight. Therefore, replication of such variances when
calibrating in the lab is essential to acquire the appropriate
sensitivities. This requirement resulted in the development of the APACHE
(All Pressure Altitude-based Calibrator for HOx Experimentation) chamber to characterize the sensitivity of the airborne LIF-FAGE (fluorescence assay by gas expansion) HOx instrument, HORUS, which took part in an intensive airborne campaign, OMO-Asia 2015. It utilizes photolysis of water vapor but has the additional ability to alter the pressure at the nozzle of the HORUS instrument. With APACHE, the HORUS instrument sensitivity towards OH (26.1–7.8 cts s−1 pptv−1 mW−1, ±22.6 % 1σ; cts stands for counts by the detector) and HO2 (21.2–8.1 cts s−1 pptv−1 mW−1, ±22.1 % 1σ) was characterized to the external pressure range at the instrument nozzle of 227–900 mbar. Measurements supported by a computational fluid dynamics model, COMSOL Multiphysics, revealed that, for all pressures explored in this study, APACHE is capable of initializing a homogenous flow and maintaining near-uniform flow speeds across the internal cross section of the chamber. This reduces the uncertainty regarding average exposure times across the mercury (Hg) UV ring lamp. Two different actinometrical approaches characterized the APACHE UV ring lamp flux as 6.37×1014(±1.3×1014) photons cm−2 s−1. One approach used the HORUS instrument as a transfer standard in conjunction with a calibrated on-ground calibration system traceable to NIST standards, which characterized the UV ring lamp flux to be 6.9(±1.1)×1014 photons cm−2 s−1. The second approach involved measuring ozone production by the UV ring lamp using an ANSYCO O3 41 M ozone monitor, which characterized the UV
ring lamp flux to be 6.11(±0.8)×1014 photons cm−2 s−1. Data presented in this study are the first direct calibrations of an airborne HOx instrument, performed in a controlled environment in the lab using APACHE.
Publisher
Copernicus GmbH
Subject
Atmospheric Science
Reference53 articles.
1. Albrecht, S. R., Novelli, A., Hofzumahaus, A., Kang, S., Baker, Y., Mentel, T., Wahner, A., and Fuchs, H.: Measurements of hydroperoxy radicals (HO2) at atmospheric concentrations using bromide chemical ionisation mass spectrometry, Atmos. Meas. Tech., 12, 891–902, https://doi.org/10.5194/amt-12-891-2019, 2019. 2. Atkinson, R., Baulch, D. L., Cox, R. A., Crowley, J. N., Hampson, R. F., Hynes, R. G., Jenkin, M. E., Rossi, M. J., and Troe, J.: Evaluated kinetic and photochemical data for atmospheric chemistry: Volume I – gas phase reactions of Ox, HOx, NOx and SOx species, Atmos. Chem. Phys., 4, 1461–1738, https://doi.org/10.5194/acp-4-1461-2004, 2004. 3. Brauers, T., Aschmutat, U., Brandenburger, U., Dorn, H. P., Hausmann, M.,
Hessling, M., Hofzumahaus, A., Holland, F., PlassDulmer, C., and Ehhalt, D.
H.: Intercomparison of tropospheric OH radical measurements by multiple folded long-path laser absorption and laser induced fluorescence,
Geophys. Res. Lett., 23, 2545–2548, https://doi.org/10.1029/96gl02204, 1996. 4. Brauers, T., Hausmann, M., Bister, A., Kraus, A., and Dorn, H. P.: OH
radicals in the boundary layer of the Atlantic Ocean 1. Measurements by
long-path laser absorption spectroscopy, J. Geophys. Res., 106, 7399–7414, https://doi.org/10.1029/2000jd900679, 2001. 5. Brune, W. H., Stevens, P. S., and Mather, J. H.: Measuring OH and HO2 in the Troposphere by Laser-Induced Fluorescence at Low-Pressure, J. Atmos. Sci., 52, 3328–3336,
https://doi.org/10.1175/1520-0469(1995)052<3328:Moahit>2.0.Co;2, 1995.
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|