Affiliation:
1. Chinese Academy of Sciences
Abstract
Calcium is generally determined by EDTA titration after separation of phosphate radicals,
and the phosphorus is determined by weigh method. This traditional analytical process is time
consuming and unfavourable for quality controlling. In contrast, the ICP-AES is simple and fast,
and can simultaneously determine multielements. In this paper, the ICP-AES method was used to
simultaneously determine calcium and phosphorous in calcium phosphate based bioceramics, and
wavelengths of 317.933 nm and 213.618 nm were selected for the measuring of the calcium and
phosphorus, respectively. The results obtained by ICP-AES are well consistent with the results
determined by the traditional EDTA titration and weigh method, and suggest that the ICP-AES
analysis is a simple, fast and accurate method for simultaneous determination of calcium and
phosphorous in calcium phosphate based bioceramics.
Publisher
Trans Tech Publications, Ltd.
Subject
Mechanical Engineering,Mechanics of Materials,General Materials Science
Reference6 articles.
1. R.Z. LeGeros, J.P. LeGeros, in: L.L. Hench, J. Wilson (Eds. ), An Introduction to Bioceramics, World Scientific, Singapore, (1993).
2. P. Ducheyne, S. Radin, L. King: J. Biomed. Mater. Res. Vol. 27 (1993), p.25.
3. S. Raynaud, E. Champion, D.B. Assollant, et al: Biomaterials Vol. 23 (2002), p.1065.
4. S. Raynaud, E. Champion, D.B. Assollant: Biomaterials Vol. 23 (2002), p.1073.
5. S. Raynaud, E. Champion, J.P. Lafon, et al: Biomaterials Vol. 23 (2002), p.1081.