Author:
Miller Brandon A.,Salehi Afshin,Limbrick David D.,Smyth Matthew D.
Abstract
OBJECTIVEThe ROSA device is a robotic stereotactic arm that uses a laser system to register the patient’s head or spine with MR or CT images. In this study, the authors analyze their experience with this system in pediatric neurosurgical applications and present selected cases that exemplify the usefulness of this system.METHODSThe authors reviewed all cases that utilized the ROSA system at their institution. Patient demographics, pathology, complications, electrode placement, laser ablation, and biopsy accuracy were analyzed. Patient disposition and condition at follow-up were also analyzed.RESULTSSeventeen patients underwent 23 procedures using the ROSA system. A total of 87 electroencephalography electrodes were placed, with 13% deviating more than 3 mm from target. Six patients underwent stereotactic needle biopsy, and 9 underwent laser interstitial thermotherapy (LITT). One patient who underwent LITT required a subsequent craniotomy for tumor resection. Another patient experienced an asymptomatic extraaxial hematoma that spontaneously resolved. No patient suffered neurological complications during follow-up. Follow-up from the last procedure averaged 180 days in epilepsy patients and 309 days in oncology patients.CONCLUSIONSThe precision, ease of use, and versatility of the ROSA system make it well suited for pediatric neurosurgical practice. Further work, including long-term analysis of results and cost-effectiveness, will help determine the utility of this system and if its applications can be expanded.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Cited by
63 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献