Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients

Author:

Butter Christian1,Meisel Eckhard1,Tebbenjohanns Juergen1,Engelmann Lothar1,Fleck Eckart1,Schubert Bernd1,Hahn Stephen1,Pfeiffer Dietrich1

Affiliation:

1. From Humboldt University and German Heart Institute, Berlin, Germany (C.B., E.F.); Heart and Circulation Center, Dresden, Germany (E.M.); Medical School, Hannover, Germany (J.T.); University Hospital, Leipzig, Germany (L.E., D.P.); and Guidant Corporation, St Paul, Minnesota, and Brussels, Belgium (B.S., S.H.).

Abstract

Background Defibrillation thresholds (DFT) with standard implantable cardioverter-defibrillator leads in the right ventricle (RV) may be determined by weak shock field intensity in the myocardium of the left ventricle (LV). Adding a shocking electrode in a coronary vein on the middle of the LV free wall, thereby establishing biventricular defibrillation, substantially reduced defibrillation requirements in animals. We investigated the feasibility of this approach in 24 patients receiving an implantable cardioverter-defibrillator using a prototype over-the-wire temporary LV defibrillation lead. Methods and Results The LV lead was inserted through the coronary sinus, using a guide catheter and guidewire, into a posterior or lateral coronary vein whose location was determined by retrograde venography. Paired DFT testing compared a standard system (RV to superior vena cava plus can emulator [SVC+Can], 60% tilt biphasic shock) to a system including the LV lead. The biventricular system was tested with a dual-shock waveform (20% tilt monophasic shock from LV→SVC+Can, then 60% tilt biphasic shock from RV→SVC+Can). Twenty patients completed DFT testing. Venography and LV lead insertion time was 46±40 minutes. The biventricular system reduced mean DFT by 45% (8.9±1.1 J versus 4.9±0.5 J, P <0.001). Twelve patients (60%) had a standard system DFT ≥8 J, and the biventricular system gave a lower DFT in all patients. There were no adverse events related to the use of the LV lead, which was removed after testing. Conclusions Internal defibrillation using a transvenously inserted LV lead is feasible, produces significantly lower DFTs, and seems safe under the conditions tested. Biventricular defibrillation may be a useful option for reducing DFTs or could be added to an LV pacing lead for heart failure.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 26 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3