Steroid Elution Improves the Stimulation Threshold in an Active-Fixation Atrial Permanent Pacing Lead

Author:

Crossley George H.1,Brinker Jeffrey A.1,Reynolds Dwight1,Spencer William1,Johnson W. Ben1,Hurd Howard1,Tonder Lisa1,Zmijewski Matt1

Affiliation:

1. From the Cardiology Section, Department of Medicine, Bowman Gray School of Medicine, Wake Forest University, Winston-Salem, NC (G.H.C.); Johns Hopkins University, Baltimore, Md (J.A.B.); The University of Oklahoma, Oklahoma City (D.R.); Baylor College of Medicine, Houston, Tex (W.S.); Mercy Hospital, Des Moines, Iowa (W.B.J.); Washington University, St Louis, Mo (H.H.); and Medtronic Inc, Minneapolis, Minn (L.T., M.Z.).

Abstract

Background Prior work suggests that the addition of a steroid-eluting reservoir to a passive-fixation permanent pacemaker lead improves the stimulation threshold; however, no large randomized study has addressed this issue. Over the last several years, there has been an increase in enthusiasm for the use of active-fixation permanent pacemaker leads for various reasons in spite of the generally accepted notion that active-fixation leads have higher stimulation thresholds. Methods and Results This multicenter, randomized, controlled study examined the difference in performance between a standard active-fixation atrial lead (Medtronic model 4058) and a steroid-eluting lead (Medtronic model 4068). Stimulation thresholds were obtained in a four-point strength-duration fashion. Evaluations of sensing and impedance were performed as well. These evaluations were performed at implantation, at weeks 1 through 4, and at weeks 6, 12, 24, and 52. Stimulation thresholds were significantly better in the steroid lead than in the nonsteroid lead at each measurement point from 1 week to 12 months. The mean 1.6-V stimulation threshold at 12 months was 0.19±0.2 ms in the steroid lead and 0.41±0.30 ms in the control lead. No acute peaking was observed with the steroid lead, whereas significant peaking was observed with the control lead. There was no difference in long-term sensing or impedance. Conclusions Inclusion of a steroid-eluting reservoir in an active-fixation permanent pacing lead improved stimulation thresholds in both the subacute and chronic periods and therefore should extend pulse-generator longevity.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3