Abstract

The goal of this study was to validate a newly improved noninvasive method for calibrated measurement of the ascending portion of the central arterial pressure wave in humans. Noninvasive pressure waveforms were generated by measuring the time delay between the R wave of the electrocardiogram and onset of brachial artery flow (by Doppler) during computer-controlled upper arm cuff deflation. This delay shortens with falling cuff pressure (becoming near constant at and below diastolic pressure), so that a plot of pressure versus time delay yields the ascending portion of the arterial waveform. These waveforms were compared with simultaneous invasive ascending aortic pressures in 57 adult patients (31 by fluid manometer [group A] and 26 by catheter-tipped micromanometer [group B]) during routine cardiac catheterization. Patient age ranged from 26 to 77 years. Eighty percent of group A patients and 40% of group B had coronary artery disease. Noninvasive systolic and diastolic pressures were very similar to invasive values in both groups (Pni = 0.98 x Pi, r = 0.99, p < 0.0001). Instantaneous pressure differences between waveforms were also similar in both groups, averaging between 4.5 and 5.5 mm Hg. Micromanometer and noninvasive pressure data were also obtained before and after intravenous nitroglycerin (n = 5) and isometric handgrip (n = 8) and demonstrated good agreement. A potential application of these pressures is for estimating maximal ventricular power to assess systolic function. This was tested using invasive pressure-volume data from four patients under a variety of conditions (exercise, pacing, etc.).(ABSTRACT TRUNCATED AT 250 WORDS)

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Internal Medicine

Reference23 articles.

1. O'Rourke MF Kelly R Avolio A: The Arterial Pulse. Philadelphia Pa Lea & Febiger 1992

2. Non-invasive registration of the arterial pressure pulse waveform using high-fidelity applanation tonometry;Kelly R;J Vase Med Biol,1989

3. Aortic input impedance in normal man: relationship to pressure wave forms.

4. Nichols WW O'Rourke MF: Contours of pressure and flow in arteries in McDonald's Blood Flow in Arteries ed 3. Philadelphia. Pa Lea & Febiger 1990 pp 143-195

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3