1. From the Divisions of Cardiology (M.I.C., D.M.B., V.L.V., R.E.T., T.S.W., L.A.R.) and Cardiothoracic Surgery (J.W.G.) and the Departments of Pediatrics (M.I.C., D.M.B., V.L.V., R.E.T., T.S.W., L.A.R.) and Surgery (J.W.G.), The Children’s Hospital of Philadelphia and the University of Pennsylvania School of Medicine, Philadelphia.

Abstract

Background —The purpose of this study was to evaluate the long-term outcome of all pediatric epicardial pacing leads. Methods and Results —All epicardial leads and 1239 outpatient visits between January 1, 1983, and June 30, 2000, were retrospectively reviewed. Pacing and sensing thresholds were reviewed at implant, at 1 month, and at subsequent 6-month intervals. Lead failure was defined as the need for replacement or abandonment due to pacing or sensing problems, lead fracture, or phrenic/muscle stimulation. A total of 123 patients underwent 207 epicardial lead (60 atrial/147 ventricular, 40% steroid) implantations (median age at implant was 4.1 years [range 1 day to 21 years]). Congenital heart disease was present in 103 (84%) of the patients. Epicardial leads were followed for 29 months (range 1 to 207 months). The 1-, 2-, and 5-year lead survival was 96%, 90%, and 74%, respectively. Compared with conventional epicardial leads, both atrial and ventricular steroid leads had better stimulation thresholds 1 month after implantation; however, only ventricular steroid leads had improved chronic pacing thresholds (at 2 years: for steroid leads, 1.9 μJ [from 0.26 to 16 μJ]; for nonsteroid leads, 4.7 μJ [from 0.6 to 25 μJ]; P <0.01). Ventricular sensing was significantly better in steroid leads 1 month after lead implantation (at 2 years: for steroid leads, 8 mV [from 4 to 31 mV]; for nonsteroid leads, 4 mV [from 0.7 to 10 mV]; P <0.01). Neither congenital heart disease, lead implantation with a concomitant cardiac operation, age or weight at implantation, nor the chamber paced was predictive of lead failure. Conclusions —Steroid epicardial leads demonstrated relatively stable acute and chronic pacing and sensing thresholds. In this evaluation of >200 epicardial leads, lead survival was good, with steroid-eluting leads demonstrating results similar to those found with historical conventional endocardial leads.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 143 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3