Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein

Author:

Bunch T. Jared1,Bruce G. Keith1,Johnson Susan B.1,Sarabanda Alvaro1,Milton Mark A.1,Packer Douglas L.1

Affiliation:

1. From the Division of Cardiovascular Disease, Department of Internal Medicine, Mayo Clinic, Rochester, Minn.

Abstract

Background— Many ablative approaches in or near the orifice of the pulmonary vein (PV) have demonstrated success in eliminating atrial fibrillation. Despite current practice, there are no data regarding the in vivo efficacy and safety of an 8-mm catheter tip for ablation at the PV orifice. Methods and Results— Ten mongrel dogs were studied. Thermocouples were implanted in the atrial muscle of the PV orifice. Intracardiac echocardiography monitored catheter position, tip/tissue orientation, and microbubble formation. Ninety-four ablations were performed for 120 seconds. A temperature discrepancy >10°C between the catheter tip and tissue occurred during 47 (50%) of the ablations. Despite termination of energy delivery, the average tissue temperature remained within 1°C of the achieved steady state for 9 seconds. A temperature discrepancy >10°C was more common in the right superior PV, with oblique catheter positioning, when tissue temperatures were >60°C or 80°C, and with type 1 or type 2 microbubble formation. However, microbubbles were not present in 7 (13%, type 1) and 10 (40%, type 2) ablations with tissue temperatures >80°C. The maximum tissue temperature achieved with non–full-thickness lesions was 47.3±7.4°C vs 75.9±11.7°C ( P <0.0001) for full-thickness lesions. Conclusions— Marked discrepancies between catheter-tip and tissue temperatures occurred with higher temperatures, prolonged ablation times, and unfavorable catheter thermistor–tissue contact. Also, these data suggest a conservative approach to atrial ablation, because full-thickness lesions were obtained when tissue temperatures reached 50°C to 60°C and the tissue retained high heat levels despite termination of radiofrequency energy. Finally, microbubbles are inconsistent markers of tissue overheating.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3