Nonuniform Strut Distribution Correlates With More Neointimal Hyperplasia After Sirolimus-Eluting Stent Implantation

Author:

Takebayashi Hideo1,Mintz Gary S.1,Carlier Stéphane G.1,Kobayashi Yoshio1,Fujii Kenichi1,Yasuda Takenori1,Costa Ricardo A.1,Moussa Issam1,Dangas George D.1,Mehran Roxana1,Lansky Alexandra J.1,Kreps Edward1,Collins Michael B.1,Colombo Antonio1,Stone Gregg W.1,Leon Martin B.1,Moses Jeffrey W.1

Affiliation:

1. From the Cardiovascular Research Foundation, Lenox Hill Heart and Vascular Institute, New York, NY.

Abstract

Background— Little is known about causes of intimal hyperplasia (IH) after sirolimus-eluting stent (SES) implantation. Methods and Results— Intravascular ultrasound was performed in 24 lesions with intra-SES restenosis and a comparison group of 25 nonrestenotic SESs. To assess stent strut distribution, the maximum interstrut angle was measured with a protractor centered on the stent, and the visible struts were counted and normalized for the number of stent cells. In SES restenosis patients, minimum lumen site was compared with image slices 2.5, 5.0, 7.5, and 10.0 mm proximal and distal to this site. The minimum lumen site had a smaller IVUS lumen area at follow-up (2.7±0.9 versus 6.2±1.9 mm 2 ; P <0.01), larger maximum interstrut angle (135±39° versus 72±23°; P <0.01), larger IH area (3.4±1.5 versus 0.6±1.1 mm 2 ; P <0.01) and thickness (0.7±0.3 versus 0.1±0.2 mm; P <0.01) at maximum interstrut angle, and fewer stent struts (4.9±1.0 versus 6.0±0.5; P <0.01) even when normalized for the number of stent cells (0.78±0.15 versus 0.97±0.07; P <0.01). Compared with nonrestenotic SES, the restenosis lesions also had a smaller minimal lumen area, larger IH area, thicker IH at maximum interstrut angle, fewer stent struts, and larger maximum interstrut angle. Multivariate analysis identified the number of visualized stent struts normalized for the number of stent cells and maximum interstrut angle as the only independent IVUS predictor of IH cross-sectional area ( P <0.01 and P <0.01), minimum lumen area ( P <0.01 and P <0.01), and IH thickness ( P <0.01 and P <0.01). Conclusions— The number and distribution of stent struts affect the amount of neointima after SES implantation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 107 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. DES Selection for Left Main and Coronary Bifurcation Stenting;Reviews in Cardiovascular Medicine;2023-09-21

2. Treatment of Ostial Right Coronary Artery Narrowings: Outcomes From the Multicenter Prospective e-ULTIMASTER Registry;Journal of the Society for Cardiovascular Angiography & Interventions;2023-05

3. Intravascular imaging in coronary stent restenosis: Prevention, characterization, and management;Frontiers in Cardiovascular Medicine;2022-08-09

4. Simulation of stent mechanical testing;Cardiovascular and Respiratory Bioengineering;2022

5. In-stent restenosis: symptoms, hemodynamic signs, pathogenesis and treatment;Kardiologicheskii vestnik;2021

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3