Performance of two plasma separation devices for HIV-1 viral load measurement in primary healthcare settings

Author:

Vubil Adolfo1ORCID,Zicai Ana Flora1,Sitoe Nádia1,Nhachigule Carina1,da Costa Paulino1,Magul Cacildo1,Meggi Bindiya1,Viegas Sofia1,Mabunda Nédio1,Jani Ilesh1

Affiliation:

1. Laboratory Reference Service, Instituto Nacional de Saúde , Marracuene, Mozambique

Abstract

ABSTRACT Dried blood spot (DBS) may overestimate the viral RNA, mainly in patients with low viral load (VL), due to proviral DNA and intracellular RNA. The Burnett and HemaSpot provide integrated solutions for the collection, separation, and drying of plasma from whole blood. This study aims to evaluate the performance of both devices compared to plasma to identify antiretroviral therapy (ART) failure. The devices were separately evaluated in a cross-sectional design. Patients on ART were included for the studies (Burnett: 611, October 2019 to January 2020) and (HemaSpot: 620, November 2020 to April 2021). VL was tested using CAP/CTM96. The sensitivity and specificity of DBS, Burnett, and HemaSpot were determined, and plasma results were considered as a reference at a threshold of 1,000 copies/ml. For the Burnett study, 2,444 specimens, including plasma, DBS, venous Burnett (vBurnett), and capillary Burnett (cBurnett), were collected. Sensitivity of DBS, vBurnett, and cBurnett was 97.4%, 98.3%, and 97.5%, respectively, whereas specificity was 86.8% for DBS, 96.9% for vBurnett, and 93.9% for cBurnett. For the HemaSpot study, 1,860 specimens were collected, including plasma, DBS, and vHemaSpot. Sensitivity of DBS and vHemaSpot was 95.0% and 91.3%, respectively, whereas specificity was 86.9% for DBS and 94.5% for vHemaSpot. The misclassification rate was more prominent in DBS (4.8%) and HemaSpot (8.4%) but lower in vBurnett (2.0%) and cBurnett (3.2%). The Burnett showed better performance than DBS, whereas HemaSpot showed poorer performance than DBS. Nevertheless, both Burnett and HemaSpot have high rate of non-reportable results. In the current format, neither of the two devices is feasible for VL scale-up in resource-limited settings. IMPORTANCE Burnett and HemaSpot are two novel technologies that allow whole blood collection and plasma separation and stabilization at room temperature without the need of additional equipment. Hence, these devices are potential alternatives to fresh plasma as a suitable specimen for viral load scale-up to monitor antiretroviral therapy in resource-limited settings

Publisher

American Society for Microbiology

Subject

Infectious Diseases,Cell Biology,Microbiology (medical),Genetics,General Immunology and Microbiology,Ecology,Physiology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3