Assay validity of point-of-care platelet function tests in thrombocytopenic blood samples

Author:

Quehenberger Peter1,Löffler Markus W.2,Wiegele Marion3,Schaden Eva3,Lacom Conrad3,Tolios Alexander4,Eichelberger Beate4

Affiliation:

1. Department of Laboratory Medicine, Medical University of Vienna, Vienna, Austria

2. Department of Immunology, Interfaculty Institute for Cell Biology, University of Tübingen, Tübingen, Germany

3. Department of Anaesthesia, Critical Care and Pain Medicine, Division of General Anaesthesia and Intensive Care Medicine, Medical University of Vienna, Vienna, Austria

4. Department of Blood Group Serology and Transfusion Medicine, Medical University of Vienna, Vienna, Austria

Abstract

Point-of-care (POC) platelet function tests are faster and easier to perform than in-depth assessment by flow cytometry. At low platelet counts, however, POC tests are prone to assess platelet function incorrectly. Lower limits of platelet count required to obtain valid test results were defined and a testing method to facilitate comparability between different tests was established. We assessed platelet function in whole blood samples of healthy volunteers at decreasing platelet counts (> 100, 80-100, 50-80, 30-50 and < 30 x109/L) using two POC tests: impedance aggregometry and in-vitro bleeding time. Flow cytometry served as the gold standard. The number of platelets needed to reach 50% of the maximum function (ED50) and the lower reference limit (EDref) were calculated to define limits of test validity. The minimal platelet count required for reliable test results was 100 x109/L for impedance aggregometry and in-vitro bleeding time but only 30 x109/L for flow cytometry. Comparison of ED50 and EDref showed significantly lower values for flow cytometry than either POC test (P value < 0.05) but no difference between POC tests nor between the used platelet agonists within a test method. Calculating the ED50 and EDref provides an effective way to compare values from different platelet function assays. Flow cytometry enables correct platelet function testing as long as platelet count is > 30 x109/L whereas impedance aggregometry and in-vitro bleeding time are inconsistent unless platelet count is > 100 x109/L.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry (medical),Clinical Biochemistry

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3