Technique for reducing the surgical porosity of small-diameter vascular grafts

Author:

Nemets E. A.1,Khairullina A. I.2,Belov V. Yu.3,Surguchenko V. A.1,Vasilets V. N.4,Sevastianov V. I.3,Volkova E. A.1,Basok Yu. B.1

Affiliation:

1. Shumakov National Medical Research Center of Transplantology and Artificial Organs

2. Moscow Institute of Physics and Technology

3. Shumakov National Medical Research Center of Transplantology and Artificial Organs; Institute of Biomedical Research and Technology

4. Semenov Federal Research Center for Chemical Physics

Abstract

High surgical porosity (SP) is one of the causes of significant blood loss, as well as hematoma formation. So, reducing the SP of small-diameter vascular grafts (VGs) is a crucial task.The objective of this work was to develop a technology for the formation of polycaprolactone (PCL)-based small-diameter VGs with a bioactive coating with reduced SP.Materials and methods. Porous VGs with an inner diameter of 3 mm were fabricated by electrospinning from 5% PCL solution with addition of 5–30% gelatin (PCL/G) on an NANON-01A unit (MECC CO, Japan). Bioactive coating was applied by sequential incubation of VGs in solutions of bovine serum albumin, heparin and platelet lysate with fixation in a glutaric aldehyde solution. The surface structure and mechanical properties of the samples were investigated. Functional properties of the bioactive VGs were evaluated in relation to their interaction with cell cultures in vitro.Results. It was found that introduction of gelatin into the working solution reduces SP from 30.4 ± 1.5 mL/(cm2 ·min) to 2.8 ± 0.5 ml/(cm2 ·min). It was shown that at a PCL/gelatin ratio of 9 : 1, the outer and inner sides of the bioactive VGs samples are characterized by surface uniformity (no defects), mechanical properties close to blood vessels of the same diameter (Young’s modulus 6.7 ± 2.1 MPa, tensile strength 26.7 ± 4.9 N and elongation to break 423 ± 80%) and ability to support adhesion and proliferation of human umbilical vein endothelial cell line, EA.hy926.Conclusion. Introduction of 10% gelatin content (by the polymer weight) into PCL solution reduces the SP of small-diameter VGs, leads to uniformity in their inner and outer surface, improvement in their mechanical properties without reducing their ability to support adhesion and proliferation of vascular endothelial cells.

Publisher

V.I. Shimakov Federal Research Center of Transplantology and Artificial Organs

Subject

Transplantation,Immunology and Allergy

Reference27 articles.

1. Blood vessel prostheses. General technical requirements. Test methods: GOST 31514-2012. Date of introduction 01.01.2015. M.: Standartinform, 2015.

2. Szentivanyi A, Chakradeo T, Zernetsch H, Glasmacher B. Electrospun cellular microenvironments: understanding controlled release and scaffold structure. Adv Drug Deliv Rev. 2011; 63: 209–220.

3. Novikova SP, Salokhedinova RR, Loseva SV, Nikolashina LN, Levkina AYu. Analysis of physico-mechanical and structural characteristics of vascular prostheses. Thoracic and cardiovascular surgery. 2012; 54: 27–33.

4. Wesolowski SA, Fries CC, Karlson KE, De Bakey M, Sawyer PN. Porosity: primary determinant of ultimate fate of synthetic vascular grafts. Surgery. 1961; 50: 91–96.

5. Lebedev LV, Plotnik LL, Smirnov AD. Protezy krovenosnykh sosudov. L.: Medicina, 1981; 192.