Abstract
ObjectiveTo assess the safety and feasibility of platelet transfusion through small-bore long lines used in the neonatal intensive care unit (NICU), including double-lumen umbilical venous catheters (UVCs) and 24 G and 28 G peripherally inserted central catheters (PICCs).DesignProspective in vitro controlled study.SettingBlood transfusion service laboratory.MethodsIn vitro platelet transfusions were set up as per NICU practice. Transfusion line pressure was monitored. Post-transfusion swirling, presence of aggregates, pH analysis and automated cell count in vitro activation response by flow cytometry assessing CD62P expression were assessed.Main outcome measuresAll transfusions completed successfully. The rate of infusion was reduced in 5 of 16 transfusions through 28 G lines due to ‘pressure high’ alarms. There was no difference in swirling values or transfusion aggregate formation, CD62P expression levels, platelet count, platelet distribution width, mean platelet volume, plateletcrit or platelet to large cell ratio across transfusions post-transfusion.ConclusionsThis study showed that in vitro platelet transfusion performed through 24 G and 28 G neonatal PICC lines and double-lumen UVCs is non-inferior to 24 G short cannulas, using outcome measures of platelet clumping, platelet activation and line occlusion. This suggests that where available these lines can be used if necessary for platelet transfusion.