Author:
Li Chenkai,Zhang Zian,Wang Guanrong,Rong Chun,Zhu Wanping,Lu Xinzhe,Liu Yikai,Zhang Haining
Abstract
Abstract
Objective
This study assessed the accuracy of robotic-arm-assisted total knee arthroplasty (RATKA) for bone resection, component size prediction, implant placement, and limb alignment.
Methods
This prospective cohort study included 36 patients. All procedures were performed by a single experienced surgeon, using an identical approach and implant designs. The MAKO RIO Robotic Interactive Orthopaedic Arm (Stryker, Mahwah, NJ, USA) system was used. The actual bone resection, implant placement, component size, and postoperative mechanical alignment were recorded, then compared with the preoperative plan.
Results
The mean absolute differences from the plan for the distal (medial and lateral) and posterior (medial and lateral) femoral cuts were 0.39 mm (0.62), 0.49 mm (0.70), 0.62 mm (0.79), and 0.65 mm (0.81), respectively, with 0.57° (0.65) varus. The mean absolute differences in the medial and lateral tibial cuts were 0.56 mm (0.75) and 0.58 mm (0.76), with 0.48° (0.16) varus and 0.54° (0.25) anterior/posterior slope. Of 192 bone resections, 176 (91.7%) were within ≤ 1 mm of the preoperative plan. The accuracies of femoral and tibial component size prediction were 100% and 97.22%, respectively. The mean absolute difference in final limb coronal alignment was 0.92° (0.65). Of the alignments, 18 (75.0%) were within ≤ 1.00° of the plan, and 100% were within ≤ 3.00° of the plan.
Conclusion
RATKA could accurately predict the component size and execute a preoperative plan to achieve precise bone resection, and implant placement, thereby reducing alignment outliers.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
19 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献