Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU

Author:

Pellet Pierre-Louis,Stevic Neven,Degivry Florian,Louis Bruno,Argaud Laurent,Guérin Claude,Cour MartinORCID

Abstract

Abstract Background Inhaled sedation during invasive mechanical ventilation in patients with acute respiratory distress syndrome (ARDS) has received increasing attention. However, inhaled sedation devices increase dead-space ventilation and an undesirable effect is the increase in minute ventilation needed to maintain CO2 removal. A consequence of raising minute ventilation is an increase in mechanical power (MP) that can promote lung injury. However, the effect of inhaled sedation devices on MP remains unknown. Methods We conducted a bench study to assess and compare the effects of three devices delivering inhaled sevoflurane currently available in ICU (AnaConDa-50 mL (ANA-50), AnaConDa-100 mL (ANA-100), and MIRUS) on MP by using a test lung model set with three compliances (20, 40, and 60 mL/cmH2O). We simulated lung-protective ventilation using a low tidal volume and two levels of positive end-expiratory pressure (5 and 15 cmH2O) under ambient temperature and dry conditions. Following the insertion of the devices, either the respiratory rate or tidal volume was increased in 15%-steps until end-tidal CO2 (EtCO2) returned to the baseline value. MP was calculated at baseline and after EtCO2 correction using a simplified equation. Results Following device insertion, the EtCO2 increase was significantly greater with MIRUS (+ 78 ± 13%) and ANA-100 (+ 100 ± 11%) than with ANA-50 (+ 49 ± 7%). After normalizing EtCO2 by adjusting minute ventilation, MP significantly increased by more than 50% with all inhaled sedation devices compared to controls. The lowest increase in MP was observed with ANA-50 (p < 0.05 versus ANA-100 and MIRUS). The Costa index, another parameter assessing the mechanical energy delivered to the lungs, calculated as driving pressure × 4 + respiratory rate, significantly increased by more than 20% in all experimental conditions. Additional experiments performed under body temperature, ambient pressure, and gas saturated with water vapor conditions, confirmed the main results with an increase in MP > 50% with all devices after normalizing EtCO2 by adjusting minute ventilation. Conclusion Inhaled sedation devices substantially increased MP in this bench model of protective ventilation, which might limit their benefits in ARDS.

Publisher

Springer Science and Business Media LLC

Subject

Critical Care and Intensive Care Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3