Comparison of the performance of two real-time fluorescent quantitative PCR kits for the detection of SARS-CoV-2 nucleic acid: a study based on large real clinical samples

Author:

Wang YitingORCID,Li XuewenORCID,Wang YifeiORCID,Tu Zheyu,Xu JianchengORCID,Pan Junqi,Zhou QiORCID

Abstract

Abstract Background The global pandemic of coronavirus disease 2019 (COVID-19) has led to the development of multiple detection kits by national manufacturers for severe acute respiratory syndrome coronavirus 2 viral nucleic acid testing. The purpose of this study is to evaluate the performance of different kits (i.e., Maccura kit and Sansure kit) in real clinical work using clinical samples, which will help with the optimization of the test kits. Method During the past three months (March–May 2022), 1399 pharyngeal swabs from suspected COVID-19 patients have been initially screened using the Maccura kit in Jilin, China, and the test results were verified using the Sansure kit. The cycle threshold (Ct) values generated by the two kits were compared at different viral load levels. Correlation and consistency of the Ct values were investigated using Spearman correlation, Deming regression, and Bland–Altman plots. The cut-off Ct values of the Maccura kit were recalculated by referencing the result of the Sansure kit as a standard. Furthermore, another 163 pharyngeal swabs from suspected COVID-19 patients were collected to verify the new cut-off values. Results As a result of the Maccura kit testing, 1192 positive cases and 207 suspected COVID-19 cases were verified. After re-examination by the Sansure kit, 1118 positive cases were confirmed. The difference between the Ct values provided by the two kits was statistically significant, except for the N gene at high viral load. The Ct values obtained from the two kits presented a linear positive correlation. The Maccura kit used new cut-off Ct values of 35.00 (ORF1ab gene) and 35.07 (N gene). Based on that, the validation pass rate for the new cut-off Ct values was 91.41%. Conclusion Since the Maccura kit is found to have false positives in actual clinical work, recalculation of the cut-off values can reduce this occurrence. In order to improve the accuracy of the testing, laboratories should use two kits for COVID-19 testing, and the adjusting and optimizing of the kits for their situation are needed.

Funder

Jiancheng Xu

Qi Zhou

Publisher

Springer Science and Business Media LLC

Subject

Infectious Diseases,Virology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3