Thick shells and medially wedged posts increase foot orthoses medial longitudinal arch stiffness: an experimental study

Author:

Pelaez Ana Sofia Tavera12,Farahpour Nader13,Griffiths Ian B.4ORCID,Moisan Gabriel15

Affiliation:

1. Department of Human Kinetics Université du Québec à Trois‐Rivières Trois‐Rivières Canada

2. Faculty of Engineering Universidad de Antioquia Medellín Colombia

3. Department of Sport Biomechanics Faculty of Sport Sciences Bu Ali Sina University Hamedan Iran

4. Sports and Exercise Medicine William Harvey Research Institute Queen Mary University of London London UK

5. Groupe de Recherche sur les Affections Neuromusculosquelettiques (GRAN) Université du Québec à Trois‐Rivières Trois‐Rivières Canada

Abstract

AbstractBackgroundFoot orthoses (FOs) are commonly prescribed devices to attenuate biomechanical deficits and improve physical function in patients with musculoskeletal disorders. It is postulated that FOs provide their effects through the production of reaction forces at the foot‐FOs interface. An important parameter to provide these reaction forces is their medial arch stiffness. Preliminary results suggest that adding extrinsic additions to FOs (e.g., rearfoot posts) increases their medial arch stiffness. A better understanding of how FOs medial arch stiffness can be modulated by changing structural factors is necessary to better customise FOs for patients. The objectives of this study were to compare FOs stiffness and force required to lower the FOs medial arch in three thicknesses and two models (with and without medially wedged forefoot‐rearfoot posts).MethodsTwo models of FOs, 3D printed in Polynylon‐11, were used: (1) without extrinsic additions (mFO), and (2) with forefoot‐rearfoot posts and a 6o medial wedge (FO6MW). For each model, three thicknesses (2.6 mm, 3.0 mm, and 3.4 mm) were manufactured. FOs were fixed to a compression plate and vertically loaded over the medial arch at a rate of 10 mm/minute. Two‐way ANOVAs and Tukey post‐hoc tests with Bonferroni corrections were used to compare medial arch stiffness and force required to lower the arch across conditions.ResultsRegardless of the differing shell thicknesses, the overall stiffness was 3.4 times greater for FO6MW compared to mFO (p < 0.001). FOs with 3.4 mm and 3.0 mm thicknesses displayed 1.3‐ and 1.1‐ times greater stiffness than FOs with a thickness of 2.6 mm. FOs with a thickness of 3.4 mm also exhibited 1.1 times greater stiffness than FOs with a thickness of 3.0 mm. Overall, the force to lower the medial arch was up to 3.3 times greater for FO6MW than mFO and thicker FOs required greater force (p < 0.001).ConclusionsAn increased medial longitudinal arch stiffness is seen in FOs following the addition of 6o medially inclined forefoot‐rearfoot posts, and when the shell is thicker. Overall, adding forefoot‐rearfoot posts to FOs is significantly more efficient than increasing shell thickness to enhance these variables should that be the therapeutic aim.

Publisher

Wiley

Subject

Orthopedics and Sports Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

全球学者库

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"全球学者库"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前全球学者库共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2023 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3