Author:
Friedrich Maximilian,Son Hyeck-Soo,Hassel Oliver,Teich Lilly,Augustin Victor Aristide,Khoramnia Ramin,Auffarth Gerd Uwe,Yildirim Timur Mert
Abstract
Abstract
Background
The aim of this study was to analyze the causes and characteristics of IOL explantation within the first year after primary implantation.
Methods
In this retrospective, cross sectional database study, a database consisting of over 2500 IOL explants sent from 199 national and international doctors over the past 10 years was analyzed. All IOLs explanted within the first year after implantation were included in this analysis. Explants with insufficient information as well as phakic and Add-on IOLs were excluded. Main outcome measures were the reason for explantation, the time between implantation and explantation, as well as IOLs’ and patients’ characteristics. Additionally, the explanted IOLs were microscopically and histologically analyzed, as required.
Results
Of all explanted IOLs from the database, 1.9% (n = 50) were explanted within the first year after implantation. The most frequent reasons for early IOL explantation were IOL dislocation (32%), visual intolerance (26%), opacification (20%), and intraoperative complications (16%). The time between implantation and explantation was the shortest in cases with intraoperative complications (1.5 ± 3.1 days), followed by IOL dislocation (90.9 ± 103.9 days), visual intolerance (98.3 ± 86.5 days), opacifications (253.5 ± 124.0 days) and other indications (249.7 ± 124.0 days). Calcification of hydrophilic IOLs was the main type of opacification (80%). Notably, seven IOLs required immediate intraoperative exchange due to an intraoperative crack in the optic or a torn off haptic.
Conclusion
Indications for early IOL explantation were IOL dislocation, visual intolerance, opacification, and intraoperative complications. Especially intraoperative damages to the IOL and early calcification show a potential for improvement of affected IOLs and implantation systems.
Funder
Medizinische Fakultät Heidelberg der Universität Heidelberg