Performance evaluation of a portable bioimpedance cardiac output monitor for measuring hemodynamic changes in athletes during a head-up tilt test

Author:

Cheung Cara H. Y.1,Khaw May L.2,Tam Victor C. W.1,Ying Michael T. C.1,Lee Shara W. Y.1ORCID

Affiliation:

1. Department of Health Technology and Informatics, Faculty of Health and Social Sciences, The Hong Kong Polytechnic University, Hong Kong Special Administrative Region, China

2. Tasmanian School of Medicine, University of Tasmania, Hobart, Tasmania

Abstract

The use of impedance cardiography to monitor physiological changes in sports is rarely reported. Using head-up tilt test, we evaluated a portable noninvasive impedance cardiography device (PhysioFlow) by comparing it with a reference Doppler monitor (USCOM). Accuracy in tracking hemodynamic changes deteriorated with higher tilt, implying a gravitational influence on its performance. Stroke volume measurements were overestimated, but the changes were underestimated. Despite its convenient physical features, the suitability of PhysioFlow for sports use is questionable.

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

全球学者库

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"全球学者库"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前全球学者库共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2023 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3