Author:
Chrastný Vladislav,Komárek Michael
Abstract
AbstractDetermination of copper using inductively coupled plasma mass spectrometry (ICP-MS) suffers from polyatomic overlays originating from Na+ and Mg2+ matrix elements due to the formation of 40Ar23Na+ and 40Ar25Mg+ in the mass-to-charge ratios of 63 and 65, respectively. The collision/reaction cell technology belongs to the most modern methods used to overcome polyatomic interferences. Gas-filled collision/reaction cell can cause an additional mass bias effect influencing analytical precision of the method. In this study, the additional mass bias effect of the hexapole collision/reaction cell ICP-MS was studied on an example of n(65Cu)/n(63Cu) isotope ratio. As a result, a method for suppressing polyatomic interference on the mass-to-charge ratio of 63 and 65 was introduced and additional mass bias of the collision/reaction cell was lowered to an acceptable level.
Publisher
Springer Science and Business Media LLC
Subject
Materials Chemistry,Industrial and Manufacturing Engineering,General Chemical Engineering,Biochemistry,General Chemistry
Cited by
42 articles.