Treatment Planning and Delivery Evaluation of Volumetric Modulated Arc Therapy for Stereotactic Body Radiotherapy of Spinal Tumours: Impact of Arc Discretization in Planning System

Author:

Marchand E. L.1,Sahgal A.12,Zhang T. J.1,Millar B. A.1,Sharpe M.1,Moseley D.1,Letourneau D.1

Affiliation:

1. Department of Radiation Oncology, Princess Margaret Hospital, University Health Network, University of Toronto, Toronto, Ontario, Canada

2. Department of Radiation Oncology, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada

Abstract

The purpose of this study was to compare single arc volumetric modulated arc therapy (VMAT) to intensity modulated radiotherapy (IMRT) for spine SBRT in terms of target coverage, organ at risk (OAR) sparing and delivery performance. VMAT plans with 91 control points (VMAT-91CP) were generated for 15 spine metastases patients previously treated with a nine-field IMRT technique. VMAT and IMRT plans were compared based on target coverage, maximum spinal cord dose, maximum plan dose and volume of normal tissue receiving 20% to 80% of the prescribed dose. Treatment delivery time and monitor units (MU) were measured to determine delivery efficiency. To assess the impact of arc discretization in the treatment planning system (TPS), the VMAT-91CP plans were modified by almost doubling the number of CPs (VMAT-181CP). Planned-to-delivered dose agreement for both techniques was assessed using two types of 3D detector arrays. VMAT-91CP target coverage was equivalent to IMRT while maintaining or improving spinal cord sparing. This was achieved without increasing the volume of normal tissue receiving low or intermediate dose levels. Planned-to-delivered dose agreement equivalent to IMRT was achieved with VMAT, but required decreasing the CP angular spacing from 4° to 2° (VMAT-181CP plans). On average, VMAT-181CP plans reduced delivery time by 53% compared to IMRT. Single-arc VMAT for spine SBRT improved delivery efficiency while maintaining target coverage and OAR sparing compared to IMRT. VMAT plans generated with a CP gantry angular spacing of 2° is recommended to avoid a discretization effect in the TPS and ensure acceptable planned-to-delivered dose agreement.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3