Author:
Kotek Gyula,Nunez-Gonzalez Laura,Vogel Mika W.,Krestin Gabriel P.,Poot Dirk H. J.,Hernandez-Tamames Juan A.
Abstract
AbstractWe present and evaluate a new insight into magnetic resonance imaging (MRI). It is based on the algebraic description of the magnetization during the transient response—including intrinsic magnetic resonance parameters such as longitudinal and transverse relaxation times (T1, T2) and proton density (PD) and experimental conditions such as radiofrequency field (B1) and constant/homogeneous magnetic field (B0) from associated scanners. We exploit the correspondence among three different elements: the signal evolution as a result of a repetitive sequence of blocks of radiofrequency excitation pulses and encoding gradients, the continuous Bloch equations and the mathematical description of a sequence as a linear system. This approach simultaneously provides, in a single measurement, all quantitative parameters of interest as well as associated system imperfections. Finally, we demonstrate the in-vivo applicability of the new concept on a clinical MRI scanner.
Publisher
Springer Science and Business Media LLC
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献