Postoperative Displacement of Deep Brain Stimulation Electrodes Related to Lead-Anchoring Technique

Author:

Contarino M. Fiorella1,Bot Maarten2,Speelman Johannes D.1,de Bie Rob M. A.1,Tijssen Marina A.13,Denys Damiaan4,Bour Lo J.1,Schuurman P. Richard2,van den Munckhof Pepijn2

Affiliation:

1. Departments of Neurology Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands

2. Departments ofNeurosurgery; and Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands

3. Departments of Psychiatry, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands

4. Department of Neurology, University Medical Centre Groningen, Groningen, the Netherlands

Abstract

Abstract BACKGROUND: Displacement of deep brain stimulation (DBS) electrodes may occur after surgery, especially due to large subdural air collections, but other factors might contribute. OBJECTIVE: To investigate factors potentially contributing to postoperative electrode displacement, in particular, different lead-anchoring techniques. METHODS: We retrospectively analyzed 55 patients (106 electrodes) with Parkinson disease, dystonia, tremor, and obsessive-compulsive disorder in whom early postoperative and long-term follow-up computed tomography (CT) was performed. Electrodes were anchored with a titanium microplate or with a commercially available plastic cap system. Two independent examiners determined the stereotactic coordinates of the deepest DBS contact on early postoperative and long-term follow-up CT. The influence of age, surgery duration, subdural air volume, use of microrecordings, fixation method, follow-up time, and side operated on first was assessed. RESULTS: Subdural air collections measured on average 4.3 ± 6.2 cm3. Three-dimensional (3-D) electrode displacement and displacement in the X, Y, and Z axes significantly correlated only with the anchoring method, with larger displacement for microplate-anchored electrodes. The average 3-D displacement for microplate-anchored electrodes was 2.3 ± 2.0 mm vs 1.5 ± 0.6 mm for electrodes anchored with the plastic cap (P = .030). Fifty percent of the microplate-anchored electrodes showed 2-mm or greater (potentially relevant) 3-D displacement vs only 25% of the plastic cap–anchored electrodes (P < .01). CONCLUSION: The commercially available plastic cap system is more efficient in preventing postoperative DBS electrode displacement than titanium microplates. A reliability analysis of the electrode fixation is warranted when alternative anchoring methods are used.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Neurology (clinical),Surgery

Cited by 28 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3