A robust self-referenced 2D nyquist ghost correction for different MRI-biomarker measurements based on multi-band interleaved EPI

Author:

Liu Xiaoxi,Chen Shihui,Cui Di,Hui Edward S.,Chan Queenie,Chen Nan-Kuei,Chang Hing-Chiu

Abstract

Purpose: To enable 2D Nyquist ghost correction for multi-band interleaved echo-planar imaging (EPI) without reference scan.Methods: 2D phase errors between positive and negative echo were directly measured from the multi-band interleaved EPI data acquired with alternating readout polarity, and then corrected by using multiplexed sensitivity encoding (MUSE) framework. The proposed method was tested on brain data acquired with and without multi-band imaging under different multi-shot factors (4, 8, and 16). In addition, the 2D phase corrections for the interleaved EPI data acquired with 2D zoomed-FOV, 2D reduced-FOV, or 3D multi-band imaging were also performed with the proposed method. The performance of our proposed method was assessed with single-to-noise ratio (SNR) and ghost-to-signal ratio (GSR), and then compared with an iterative 2D phase correction method. The g-factor penalty of the proposed 2D Nyquist ghost correction method associated with 2D phase error was also evaluated. The feasibility of the proposed method was evaluated in diffusion MRI and multi-echo fMRI where quantitative biomarkers with high quality are desired.Results: The proposed method successfully suppressed the ghost artifacts for the multi-shot interleaved EPI data acquired with different multi-shot factors, 2D and 3D multi-band acquisitions, zoomed-FOV, or reduced-FOV. Compared with the iterative method, our proposed method showed lower GSR and comparable SNR performance. Through g-factor penalty simulation, the proposed method showed less than 9% of overall SNR loss associated with a maximum y-linear phase error of π. The residual aliasing artifact was effectively eliminated in the quantitative biomarkers (ADC, FA and dynamic T2*) when applying the proposed method in diffusion imaging and multi-echo fMRI.Conclusion: The proposed method can perform robust 2D Nyquist ghost correction for multi-band multi-shot interleaved EPI without reference scan, and thus can be a generalized 2D ghost correction method for multi-shot interleaved EPI based acquisitions and applications.

Publisher

Frontiers Media SA

Subject

Physical and Theoretical Chemistry,General Physics and Astronomy,Mathematical Physics,Materials Science (miscellaneous),Biophysics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3