Efficacy and Safety of Acetabular Cup without Screw Fixation in Total Hip Arthroplasty: A Systematic Review and Meta-Analysis

Author:

Miura TakanoriORCID,Kijima HiroakiORCID,Kimura RyotaORCID,Watanabe JunORCID,Okazaki Yuji,Miyakoshi Naohisa

Abstract

Background and Objectives: Adequate initial fixation of the uncemented acetabular component in total hip arthroplasty is necessary to achieve long-term survival. Although screw fixation contributes to improved cup stability, there is currently no consensus on the use of this method. This study aimed to assess the existing randomized controlled trials (RCTs) on the efficacy and safety of cup fixation in total hip arthroplasty without screws. Materials and Methods: We searched the EMBASE, MEDLINE, Cochrane Central Register of Controlled Trials, World Health Organization International Clinical Trials Registry Platform, and ClinicalTrials.gov databases to identify RCTs published before February 2022. Primary outcomes were reoperation, cup migration, and Harris Hip Score. Secondary outcomes were the presence of a radiolucent line in the acetabular region, translation and rotation movement, and polyethylene wear. We conducted meta-analyses using the random-effects models. The revised Cochrane risk-of-bias tool was used to assess the risk of bias for outcomes of interest; the Grading of Recommendations, Assessment, Development, and Evaluation approach was used to summarize the body of evidence. Results: We included six reports from four studies. Total hip arthroplasty without screw fixation to the acetabular cup had little to no effect on reoperation (pooled relative risk, 0.98; 95% confidence interval, 0.14–6.68; I2 = 0%), cup migration (pooled relative risk, 1.72; 95% confidence interval, 0.29–10.33; I2 = 1%), Harris Hip Score (mean difference, 1.19; 95% confidence interval, −1.31–3.70; I2 = 0%), radiolucent line (pooled relative risk, 5.91; 95% confidence interval, 0.32–109.35), translation and rotation of all axes, and polyethylene wear (mean difference, 0.01; 95% confidence interval, −0.01–0.04; I2 = 0%), with very low certainty of evidence on all measures. Conclusions: The efficacy of acetabular cups without screw fixation in total hip arthroplasty remains uncertain, suggesting the need for prudent clinical application. Further large-scale, well-designed studies with low risk of bias are required.

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3