Improving Sonication Efficiency in Transcranial MR-Guided Focused Ultrasound Treatment: A Patient-Data Simulation Study

Author:

Kim Changsoo1,Eames Matthew23,Paeng Dong-Guk24ORCID

Affiliation:

1. Research Institute for Basic Sciences, Jeju National University, Jeju 63243, Republic of Korea

2. Focused Ultrasound Foundation, Charlottesville, VA 22903, USA

3. Department of Radiology, University of Virginia, Charlottesville, VA 22903, USA

4. Ocean System Engineering, Jeju National University, Jeju 63243, Republic of Korea

Abstract

The potential improvement in sonication efficiency achieved by tilting the focused ultrasound (FUS) transducer of the transcranial MR-guided FUS system is presented. A total of 56 cases of patient treatment data were used. The relative position of the clinical FUS transducer to the patient’s head was reconstructed, and region-specific skull density and porosity were calculated based on the patient’s CT volume image. The total transmission coefficient of acoustic waves emitted from each channel was calculated. Then, the total energy penetrating the human skull—which represents the sonication efficiency—was estimated. As a result, improved sonication efficiency was by titling the FUS transducer to a more appropriate angle achieved in all 56 treatment cases. This simulation result suggests the potential improvement in transcranial-focused ultrasound treatment by simply adjusting the transducer angle.

Funder

National Research Foundation of Korea

Publisher

MDPI AG

Subject

Bioengineering

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3