Author:
Xu Zhanpeng,Jiang Yiming,He Sailing
Abstract
In this work, we develop a multi-mode microscopic hyperspectral imager (MMHI) for the detection of biological samples in transmission imaging, reflection imaging and fluorescence mode. A hyperspectral image cube can be obtained with 5 μm spatial resolution and 3 nm spectral resolution through push-broom line scanning. To avoid possible shadows produced by the high magnification objective with a short working distance, two illumination patterns are designed to ensure the co-axiality of the illumination and detection. Three experiments for the detection of zebrafish and fingerprints and the classification of disaster-causing microalgae verify the good capability and functionality of the system. Based on the detected spectra, we can observe the impacts of β-carotene and melanin in zebrafish, hemoglobin in the fingertip, and chlorophyll in microalgae, respectively. Multi-modes can be switched freely according to the application requirement and characteristics of different samples, like transmission mode for the transparent/translucent sample, reflection mode for the opaque sample and fluorescence mode for the fluorescent sample. The MMHI system also has strong potential for the non-invasive and high-speed sensing of bio or clinical samples.
Funder
National Key Research and Development Program of China
National Natural Science Foundation of China
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
22 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献