The Quantitative Detection of Cystatin-C in Patient Samples Using a Colorimetric Lateral Flow Immunoassay

Author:

Bikkarolla Santosh Kumar1,Venkatesan Kavipriya1,Revathy Yeddula Rebecca2,Parameswaran Sowmya2ORCID,Krishnakumar Subramanian2ORCID,Dendukuri Dhananjaya1

Affiliation:

1. Achira Labs, 66b, 13th Cross Rd, Dollar Layout, 3rd Phase, J. P. Nagar, Bengaluru 560078, India

2. Vision Research Foundation, Chennai 600006, India

Abstract

A colloidal gold-based lateral flow immunoassay was developed for the rapid quantitative detection of Cystatin-C in serum and whole blood. This device has an assay time of 15 min, making it a convenient point-of-care diagnostic tool. The device has a quantification range spanning from 0.5 to 7.5 µg/mL, with a lower limit of detection at 0.18 µg/mL. To validate its accuracy, the test was compared to a standard nephelometric immunoassay, and the results exhibited a robust linear correlation with an adjusted r2 value of 0.95. Furthermore, the device demonstrates satisfactory levels of analytical performance in terms of precision, sensitivity, and interference, indicating its potential for precise Cystatin-C quantification, particularly in renal-failure patients. Notably, the Cystatin-C-LFA device also demonstrates satisfactory stability, as a 30-day accelerated stability study at 50 °C showed no change in the device performance, indicating a long shelf life for the product when stored at room temperature.

Funder

Achira Labs. Pvt. Ltd.

Publisher

MDPI AG

Subject

Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)

Reference32 articles.

1. Initial Estimated Glomerular Filtration Rate Decline and Long-Term Renal Function During Intensive Antihypertensive Therapy;Collard;Hypertension,2020

2. The Glomerular Filtration Rate: An Important Test for Diagnosis, Staging, and Treatment of Chronic Kidney Disease;Woodhouse;Lab. Med.,2006

3. Serum Creatinine and Renal Function;Levey;Annu. Rev. Med.,1988

4. Benefits of Direct Glomerular Filtration Rate (GFR) Determination versus Creatinine-Based Tests for Evaluating Renal Function;Cimmino;J. Am. Osteopath. Assoc.,1998

5. Limitations of Creatinine as a Filtration Marker in Glomerulopathic Patients;Shemesh;Kidney Int.,1985