The Impact of Dual and Triple Energy Window Scatter Correction on I-123 Postsurgical Thyroid SPECT/CT Imaging Using a Phantom with Small Sizes of Thyroid Remnants

Author:

Michael Konstantinos12ORCID,Frangos Savvas3ORCID,Iakovou Ioannis4ORCID,Lontos Antonis15,Demosthenous George1,Parpottas Yiannis15ORCID

Affiliation:

1. Department of Mechanical Engineering, Frederick University, Nicosia 1036, Cyprus

2. Department of Medical Physics, Bank of Cyprus Oncology Center, Nicosia 2006, Cyprus

3. Department of Nuclear Medicine, Bank of Cyprus Oncology Center, Nicosia 2006, Cyprus

4. School of Medicine, Aristotle University of Thessaloniki, 54621 Thessaloniki, Greece

5. Frederick Research Center, Nicosia 1036, Cyprus

Abstract

I-123 is preferential over I-131 for diagnostic SPECT imaging after a thyroidectomy to determine the presence and size of residual thyroid tissue for radioiodine ablation. Scattering degrades the quality of I-123 SPECT images, primarily due to the penetration of high-energy photons into the main photopeak. The objective of this study was to quantitatively and qualitatively investigate the impact of two widely used window-based scatter correction techniques, the dual energy window (DEW) and triple energy window (TEW) techniques, in I-123 postsurgical SPECT/CT thyroid imaging using an anthropomorphic phantom with small sizes of remnants and anatomically correct surrounding structures. For this purpose, non-scatter-corrected, DEW and TEW scatter-corrected SPECT/CT acquisitions were performed for 0.5–10 mL remnants within a phantom, with 0.5–12.6 MBq administered activities within the remnants, and without and with background-to-remnant activity ratios of 5% and 10%. The decrease in photons, the noise and non-uniformity in the background region due to scatter correction were measured, as well as the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) from small remnants. The images were also visually evaluated by two experienced nuclear medicine physicians. Scatter correction decreased photons to a higher extent in larger regions than smaller regions. Larger remnants yielded higher SNR and CNR values, particularly at lower background activities. It was found from the quantitative analysis and the qualitative evaluation that TEW scatter correction performed better than DEW scatter correction, particularly at higher background activities, while no significant differences were reported at lower background activities. Scatter correction should be applied in I-123 postsurgical SPECT/CT imaging to improve the image contrast and detectability of small remnants within the background.

Funder

European Regional Development Fund and the Republic of Cyprus

Publisher

MDPI AG

Subject

Paleontology,Space and Planetary Science,General Biochemistry, Genetics and Molecular Biology,Ecology, Evolution, Behavior and Systematics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3