Comparison of baseline drifts using three reflector blocks versus using a single reflector block for the calibration of wall‐mounted Respiratory Gating for Scanner (RGSC) camera integrated with a CT

Author:

Liu Bei1,Shi Chengyu1,Prakash Maneesha1,Gonzalez Bryan1,Kassardjian Ari1,Kim Ji1,Mandelin Paul1,Williams Terence1,Liu An1

Affiliation:

1. Division of Radiation Oncology City of Hope National Medical Center Duarte California USA

Abstract

AbstractBackgroundThe calibration of the Respiratory Gating for SCanner (RGSC) system is critical to achieve better and more stable accuracy. The current procedure for a wall‐mounted RGSC system has a relatively large residual error.PurposeTo compare the baseline drifts in the image acquisition of DIBH using three reflector blocks versus using a single reflector block in the calibration of a wall‐mounted RGSC camera system.Materials and methodsVarian provides a calibration plate with three rows of calibration points: each row is separated by 15 cm longitudinally and by 10 cm laterally. In Varian's single‐block calibration method, the reflector block was first placed on the center point of the calibration plate and aligned with the scanner isocenter. The calibration took a picture of the block, then placed the block on the other eight points sequentially. In the proposed three‐block method, we placed three reflector blocks on the center row, with the center block aligned with the isocenter, and we took a picture of the center block by manually blocking the other two blocks in calibration. By moving the couch longitudinally in or out 15 cm, the calibration goes through all nine points. Monte Carlo simulation was done using Matlab to analyze the calibration matrix eigenvalue characteristics.ResultsFor a typical scan length of 40 cm of DIBH, the residual baseline drift in simulated DIBH is 0.02 ± 0.03  versus 0.30 ± 0.12 cm for three‐block calibration and single‐block calibration, respectively. To achieve 0.5 mm tolerance for the eigenvalue, the laser and reflector box should be within ±3 mm uncertainties based on the eigenvalue simulation.ConclusionThree‐block calibration method effectively removes baseline drift caused by couch movement in DIBH/4D CT scan for the wall‐mounted camera while the single‐block calibration method still has significant residual baseline drift.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Resolving signal drift in the wall‐mounted camera of the RGSC system;Journal of Applied Clinical Medical Physics;2024-02-02

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3