Quantitative performance of photon‐counting CT at low dose: Virtual monochromatic imaging and iodine quantification

Author:

Vrbaski Stevan123,Bache Steve4,Rajagopal Jayasai15,Samei Ehsan14

Affiliation:

1. Department of Radiology Carl E. Ravin Advanced Imaging Laboratories Duke University Medical Center Durham North Carolina USA

2. Department of Physics University of Trieste Trieste Italy

3. Elettra‐Sincrotrone Trieste, Basovizza Trieste Italy

4. Clinical Imaging Physics Group Department of Radiology Duke University Medical Center Durham North Carolina USA

5. Radiology and Imaging Sciences Clinical Center National Institutes of Health Bethesda Maryland USA

Abstract

AbstractBackgroundQuantitative imaging techniques, such as virtual monochromatic imaging (VMI) and iodine quantification (IQ), have proven valuable diagnostic methods in several specific clinical tasks such as tumor and tissue differentiation. Recently, a new generation of computed tomography (CT) scanners equipped with photon‐counting detectors (PCD) has reached clinical status.PurposeThis work aimed to investigate the performance of a new photon‐counting CT (PC‐CT) in low‐dose quantitative imaging tasks, comparing it to an earlier generation CT scanner with an energy‐integrating detector dual‐energy CT (DE‐CT). The accuracy and precision of the quantification across size, dose, material types (including low and high iodine concentrations), displacement from iso‐center, and solvent (tissue background) composition were explored.MethodsQuantitative analysis was performed on two clinical scanners, Siemens SOMATOM Force and NAEOTOM Alpha using a multi‐energy phantom with plastic inserts mimicking different iodine concentrations and tissue types. The tube configurations in the dual‐energy scanner were 80/150Sn kVp and 100/150Sn kVp, while for PC‐CT both tube voltages were set to either 120 or 140 kVp with photon‐counting energy thresholds set at 20/65 or 20/70 keV. The statistical significance of patient‐related parameters in quantitative measurements was examined using ANOVA and pairwise comparison with the posthoc Tukey honest significance test. Scanner bias was assessed in both quantitative tasks for relevant patient‐specific parameters.ResultsThe accuracy of IQ and VMI in the PC‐CT was comparable between standard and low radiation doses (p < 0.01). The patient size and tissue type significantly affect the accuracy of both quantitative imaging tasks in both scanners. The PC‐CT scanner outperforms the DE‐CT scanner in the IQ task in all cases. Iodine quantification bias in the PC‐CT (−0.9 ± 0.15 mg/mL) at low doses in our study was comparable to that of DE‐CT (range ‐2.6 to 1.5 mg/mL, published elsewhere) at a 1.7× higher dose, but the dose reduction severely biased DE‐CT (4.72 ± 0.22 mg/mL). The accuracy in Hounsfield units (HU) estimation was comparable for 70 and 100 keV virtual imaging between scanners, but PC‐CT was significantly underestimating virtual 40 keV HU values of dense materials in the phantom representing the extremely obese population.ConclusionsThe statistical analysis of our measurements reveals better IQ at lower radiation doses using new PC‐CT. Although VMI performance was mostly comparable between the scanners, the DE‐CT scanner quantitatively outperformed PC‐CT when estimating HU values in the specific case of very large phantoms and dense materials, benefiting from increased X‐ray tube potentials.

Funder

Foundation for the National Institutes of Health

Instituto Nazionale di Fisica Nucleare

Publisher

Wiley

Subject

General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3