Technical note: Angular dependence in fiber optic dosimetry using YVO4:Eu3+

Author:

Martínez Nahuel12,Fernández Yohanna1,Machello Soledad3,Molina Pablo1,Massa José4,Santiago Martín1

Affiliation:

1. CIFICEN (UNCPBA‐CICPBA‐CONICET) Tandil Argentina

2. LBN (UNQ‐CONICET) Bernal Argentina

3. CIO Tandil Argentina

4. INTIA (FCEx‐UNCPBA) Tandil Argentina

Abstract

AbstractBackgroundFiber optic dosimetry (FOD) has emerged as a useful technique that can be used in those cases when intracavitary, real time, high spatial resolution dose assessment is required. Among the several factors characterizing a dosimeter, angular response of FOD probes has to be assessed in order to consider possible clinical application.PurposeThe objective of this study was to characterize the angular response of a FOD probe based on a cylindrical shaped YVO4:Eu3+ scintillator under irradiation with a 6 MV photon beam generated by a linear accelerator (LINAC).MethodsA FOD probe was irradiated inside a plastic phantom using a 6 MV LINAC photon beam at different azimuthal angles (0° to 360°, 15° steps). Scintillation output was measured with a photomultiplier tube. Similar measurements were performed with a second FOD probe having an optical filter interposed between the scintillator and the fiber. Monte Carlo simulations using PENELOPE were carried out in order to interpret the observed results.ResultsThe FOD output was symmetrical with respect to the scintillator axis. For the unfiltered probe, the signal was maximum at rear incidence (0°) and steadily decreased down to its minimum value at frontal incidence (180°) having a signal ratio of 37%. The output of the filtered probe showed a plateau from 15° up to 115°. The signal was maximum at 60° and minimum at 180° having a signal ratio of 16%. Monte Carlo simulations predicted symmetry of the deposited dose about 0° and 90°, which contrasts with experimental findings.ConclusionsPhotoluminescence (PL) of the scintillator induced by the Cherenkov light increases the angular dependence. Radiation attenuation inside the scintillator and partial light collection of the scintillation yield by the optical fiber (OF) are responsible for asymmetrical response. Results from this study should be considered in order to minimize angular dependence in FOD.

Funder

Agencia Nacional de Promoción Científica y Tecnológica

Consejo Nacional de Investigaciones Científicas y Técnicas

Publisher

Wiley

Subject

General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3